Roche Diagnostics Corp.: Roche Online TDM Tobramycin, for use on Roche/Hitachi analyzers 917 and Modular P; catalog no. 0464250319011190. (Note: Not distributed within the United States.)

Recall number
Z-0918-2007
Recalling firm
Roche Diagnostics Corp.
Product
Roche Online TDM Tobramycin, for use on Roche/Hitachi analyzers 917 and Modular P; catalog no. 0464250319011190. (Note: Not distributed within the United States.)
Status
Terminated
Initiated
2007-04-05
Posted
2007-06-09
Terminated
2008-04-30
Reason
Reported results may have a positive bias of up to 113% because inadequate cleaning instructions may result in carry-over of the following reagents: Online TDM carbamazepine, Online TDM gentamycin, Online TDM phenytoin, Online DAT cocaine II, Online DAT methadone II, Online DAT opiates II and Online DAT THC II.
Root cause
Other
Action
Foreign affiliates were notified via email letter dated 4/5/07 requesting that to notify their customers of the need to implement a special wash cycle to prevent reagent carry over.
Quantity
474 total (Z-0917-0918-2007)
Distribution
International Distribution ONLY-Australia, Canada and Germany.
Lot, serial or model codes
Lots 144114, exp. 9/30/2007; 144167, exp. 12/31/2007; 144178, exp. 3/31/2007 and 144196, exp. 5/31/2007.
510(k) numbers
["K060853"]
PMA numbers
not on file
Product code
KLB
Firm FEI
1823260
Firm city
Indianapolis
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register