FDA Medical Device Recalls (openFDA)
MRL, Inc., A Welch Allyn Company: Welch Allyn AED 10 Defibrillator; a battery powered automated external defibrillator; Manufacturer: MRL, Inc., A Welch Allyn Company, 100 Asbury Drive, Buffalo Grove, Illinois 60089; Model AED10, part numbers 970300 (defibrillator) and 970302 (defibrillator with soft shell carrying case)
- Recall number
- Z-0918-04
- Recalling firm
- MRL, Inc., A Welch Allyn Company
- Product
- Welch Allyn AED 10 Defibrillator; a battery powered automated external defibrillator; Manufacturer: MRL, Inc., A Welch Allyn Company, 100 Asbury Drive, Buffalo Grove, Illinois 60089; Model AED10, part numbers 970300 (defibrillator) and 970302 (defibrillator with soft shell carrying case)
- Status
- Terminated
- Initiated
- 2004-04-26
- Posted
- 2004-07-20
- Terminated
- 2004-11-15
- Reason
- A faulty component has been identified which could result in an intermittent shock button (button #3) failure on the AED10 defibrillator.
- Root cause
- Other
- Action
- Urgent Medical Device Recall letters dated 4/26/04 were sent to the end users via certified mail on 4/26/04. The letters informed the users of the intemittent shock (fire or 3) button failure on the AED 10. The letters included a list of the affected serial numbers, and requested the users to call Welch Allyn MRL Technical Support as soon as possible at 1-800-462-0777 for a replacement AED 10 and to answer any questions. A replacement AED 10 will be sent to the end users within 10 days, and the users were requested to return their AED 10 in the same shipping box accompanying the replacement unit. A follow-up letter will accompany the replacement unit with instructions on how to verify the affected units, packing instructions for the return of the affected units to Welch Allyn MRL, and a correction response form to be completed and enclosed with the returned unit.
- Quantity
- 460 units
- Distribution
- Nationwide, and internationally to the United Kingdom, Ireland, Singapore, France and South Africa.
- Lot, serial or model codes
- part numbers 970300 and 970302; serial numbers 4613, 4888/4891, 4897, 6369/6386, 6396/6704, 6706, 6709/6743, 6745/6803, 6806, 6807, 6809, 6811, 6813/6832, 6836/6838, 6853/6855, 6912, 6914, 6917, 6921, 6922, 6924/6926
- 510(k) numbers
- ["K021168"]
- PMA numbers
- not on file
- Product code
- MKJ--
- Firm FEI
- 1418729
- Firm city
- Buffalo Grove
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28