FDA Medical Device Recalls (openFDA)
Stryker Sustainability Solutions: Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134301, Sterile, Rx. The Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. They are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed 2515 NAV eco Variable EP Catheters provide location information when used with compatible CARTO EP Navigation Systems version 2.3 or higher.
- Recall number
- Z-0916-2018
- Recalling firm
- Stryker Sustainability Solutions
- Product
- Stryker Sustainability Solutions Reprocessed BW Lasso 2515 NAV eco Variable Diagnostic EP Catheters, Ref D134301, Sterile, Rx. The Reprocessed 2515 NAV eco Variable Electrophysiology (EP) Catheters are indicated for multiple electrode electrophysiological mapping of the cardiac structures of the heart, i.e. recording or stimulation only. They are designed to obtain electrograms in the atrial regions of the heart. The Reprocessed 2515 NAV eco Variable EP Catheters provide location information when used with compatible CARTO EP Navigation Systems version 2.3 or higher.
- Status
- Terminated
- Initiated
- 2018-01-02
- Posted
- not on file
- Terminated
- 2018-05-02
- Reason
- An EEPROM chip error code may occur when the catheters are used with CARTO EP Navigation Systems.
- Root cause
- Device Design
- Action
- Stryker sent an Urgent Medical Device Recall letter dated January 2, 2018, explains the reason for recall, risk to health, and instructs them to discontinue use of the affected products. The customer is to check their inventory for the affected serial numbers provided in the attachment and complete the recall effectiveness check form to indicate the amount of product in inventory. Return instructions are provided. For further questions, please call (888) 763-8803.
- Quantity
- 170 devices
- Distribution
- USA (nationwide) Distribution was made to medical facilities in AZ, CA, CT, FL, GA, IL, KS, LA, MN, MO, OH, PA, TX, VT, WA, and Wi. Foreign distribution was made to Canada.
- Lot, serial or model codes
- Serial numbers 2451387, 2484264, 2484471, 2486823, 2487380, 2488350, 2490592, 2490605, 2492930, 2494064, 2494067, 2495242, 2495335, 2495337, 2495339, 2500403, 2500911, 2510524, 2518512, 2523848, 2523849, 2523851, 2525678, 2528071, 2534854, 2534856, 2608429, 2609031, 2609102, 2609105, 2609106, 2609107, 2609985, 2610794, 2610802, 2614613, 2614847, 2614909, 2615094, 2615098, 2615099, 2615319, 2615646, 2615712, 2615722, 2615725, 2615937, 2615938, 2615940, 2615943, 2615952, 2615966, 2616087, 2616088, 2616089, 2616090, 2616212, 2617794, 2617920, 2617987, 2617999, 2618010, 2618258, 2618832, 2618837, 2618881, 2618884, 2618990, 2618992, 2618993, 2619615, 2619742, 2619744, 2619827, 2620937, 2620942, 2620945, 2621283, 2621384, 2621385, 2621387, 2621424, 2621425, 2621847, 2621860, 2621880, 2623972, 2623974, 2624006, 2624010, 2624113, 2624138, 2625427, 2625536, 2625550, 2625623, 2625625, 2626330, 2626348, 2626929, 2626966, 2626988, 2627596, 2627946, 2627948, 2627949, 2629162, 2629164, 2630084, 2630085, 2630087, 2630351, 2631186, 2631188, 2631189, 2631191, 2631680, 2632641, 2633968, 2634008, 2634252, 2634271, 2634596, 2634709, 2634815, 2634836, 2635075, 2635077, 2635342, 2635343, 2635867, 2635873, 2636658, 2636759, 2636906, 2636986, 2637009, 2637022, 2637023, 2637233, 2637234, 2637237, 2637239, 2637257, 2637308, 2637495, 2637497, 2640103, 2640106, 2640107, 2640118, 2640119, 2642853, 2643222, 2643426, 2643432, 2646002, 2647122, 2669947, 2693960, 2696255, 2705129, 2708915, 2708916, 2712736, 2712756, 2713556, 2713593, 2713595, and 2713596.
- 510(k) numbers
- ["K112292"]
- PMA numbers
- not on file
- Product code
- NLH
- Firm FEI
- 3009350690
- Firm city
- Tempe
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28