Boston Scientific Corporation: Lotus TAVR 25mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV250, Catalog Number LTV25; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.
- Recall number
- Z-0916-2015
- Recalling firm
- Boston Scientific Corporation
- Product
- Lotus TAVR 25mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV250, Catalog Number LTV25; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.
- Status
- Terminated
- Initiated
- 2014-11-19
- Posted
- 2015-01-13
- Terminated
- 2015-03-26
- Reason
- Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to convert the patient to surgery.
- Root cause
- Device Design
- Action
- Boston Scientific notified consignees on November 19, 2014, of the problem via letter through overnight delivery directing consignees to segregate and return affected product to Boston Scientific utilizing the Field Safety Corrective Action (FSCA) Instructions provided. The FSCA details that consignees should: Immediately discontinue use of and segregate product; Complete and return the Account Reply Verification Tracking form and Package/Ship the affected product.
- Quantity
- 278 units - total all sizes
- Distribution
- Distributed only in the countries of Finland, France, Germany, Great Britain, Italy, Norway, Spain, Sweden, and Switzerland.
- Lot, serial or model codes
- Serial numbers: 14126051, 14126052, 14126053, 14127058, 14128047, 14128048, 14128049, 14128050, 14129004, 14129005, 14129006, 14129007, 14129008, 14133006, 14133007, 14133008, 14133009, 14133010, 14134042, 14134043, 14134044, 14134045, 14134047, 14134049, 14134050, 14134054, 14148005, 14148006, 14148007, 14148008, 14148009, 14148010, 14148011, 14148012, 14148013, 14148014, 14148015, 14148019, 14148020, 14148021, 14148022, 14148023, 14148024, 14148026, 14148027, 14156025, 14156026, 14157007, 14157008, 14157009, 14157010, 14157011, 14157012, 14157013, 14160009, 14160011, 14160012, 14160015, 14161036, 14161037, 14161038, 14161040, 14162050, 14162051, 14162053, 14162054, 14163052, 14163053, 14163054, 14163055, 14165004, 14165005, 14165006, 14165007, 14165008, 14167012, 14167013, 14167015, 14167016, 14167017, 14167018, 14167019, 14167020, 14167021, 14170001, 14170002, 14170003, 14170004, 14170005, 14170031, 14170032, 14170034, 14170035, 14170036, 14170037, 14170038, 14170039, 14170040, 14171031, 14171032, 14171033, 14171036, 14171037, 14171038, 14171039, 14171040, 14172017, 14172018, 14172019, 14172020, 14172021, 14172022, 14172023, 14172024, 14176021, 14176022, 14176023, 14176024, 14176025, 14176029, 14176030, 14176031, 14176032, 14176033, 14178036, 14178038, 14178039, 14178040, 14178041, 14178042, 14178043, 14178045, 14179048, 14179050, 14179051, 14179052, 14179053, 14179054, 14179055, 14179057, 14182027, 14182029, 14182031, 14183011, 14183013, 14183014, 14183015, 14184003, 14184004, 14184005, 14184006, 14184007, 14184008, 14184009, 14184010, 14184011, 14188009, 14188010, 14188012, 14188013, 14188014, 14188015, 14188016, 14188017, 14188018, 14191021, 14191022, 14191023, 14191024, 14191030, 14193002, 14193003, 14193007, 14193011, 14193012, 14193013, 14193014, 14193022, 14197042, 14197047, 14197048, 14197049. Expiry Datres Nov. 15, 2014 to Jan. 24, 2015.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- NPT
- Firm FEI
- 3003882783
- Firm city
- Los Gatos
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28