Skip to the content

FDA Medical Device Recalls (openFDA)

Boston Scientific Corporation: Lotus TAVR 25mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV250, Catalog Number LTV25; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.

Recall number
Z-0916-2015
Recalling firm
Boston Scientific Corporation
Product
Lotus TAVR 25mm, Transcatheter Aortic Valve Prosthesis Premounted on Delivery System; Sterile Material number H749LTV250, Catalog Number LTV25; Product Usage: The Lotus Valve System is intended to improve aortic valve function for symptomatic subjects with severe calcific aortic stenosis (aortic valve area [AVA] of <1.0 cm2 or AVA index of <0.6 em /m2) who are at extreme or high risk for standard surgical valve replacement.
Status
Terminated
Initiated
2014-11-19
Posted
2015-01-13
Terminated
2015-03-26
Reason
Lotus valve became unlocked during release from the delivery system. This may lead to percutaneous insertion of another valve, resulting in a Valve in Valve (ViV) implantation. If this occurs, the first valve could embolize and it may be necessary to convert the patient to surgery.
Root cause
Device Design
Action
Boston Scientific notified consignees on November 19, 2014, of the problem via letter through overnight delivery directing consignees to segregate and return affected product to Boston Scientific utilizing the Field Safety Corrective Action (FSCA) Instructions provided. The FSCA details that consignees should: Immediately discontinue use of and segregate product; Complete and return the Account Reply Verification Tracking form and Package/Ship the affected product.
Quantity
278 units - total all sizes
Distribution
Distributed only in the countries of Finland, France, Germany, Great Britain, Italy, Norway, Spain, Sweden, and Switzerland.
Lot, serial or model codes
Serial numbers: 14126051, 14126052, 14126053, 14127058, 14128047, 14128048, 14128049, 14128050, 14129004, 14129005, 14129006, 14129007, 14129008, 14133006, 14133007, 14133008, 14133009, 14133010, 14134042, 14134043, 14134044, 14134045, 14134047, 14134049, 14134050, 14134054, 14148005, 14148006, 14148007, 14148008, 14148009, 14148010, 14148011, 14148012, 14148013, 14148014, 14148015, 14148019, 14148020, 14148021, 14148022, 14148023, 14148024, 14148026, 14148027, 14156025, 14156026, 14157007, 14157008, 14157009, 14157010, 14157011, 14157012, 14157013, 14160009, 14160011, 14160012, 14160015, 14161036, 14161037, 14161038, 14161040, 14162050, 14162051, 14162053, 14162054, 14163052, 14163053, 14163054, 14163055, 14165004, 14165005, 14165006, 14165007, 14165008, 14167012, 14167013, 14167015, 14167016, 14167017, 14167018, 14167019, 14167020, 14167021, 14170001, 14170002, 14170003, 14170004, 14170005, 14170031, 14170032, 14170034, 14170035, 14170036, 14170037, 14170038, 14170039, 14170040, 14171031, 14171032, 14171033, 14171036, 14171037, 14171038, 14171039, 14171040, 14172017, 14172018, 14172019, 14172020, 14172021, 14172022, 14172023, 14172024, 14176021, 14176022, 14176023, 14176024, 14176025, 14176029, 14176030, 14176031, 14176032, 14176033, 14178036, 14178038, 14178039, 14178040, 14178041, 14178042, 14178043, 14178045, 14179048, 14179050, 14179051, 14179052, 14179053, 14179054, 14179055, 14179057, 14182027, 14182029, 14182031, 14183011, 14183013, 14183014, 14183015, 14184003, 14184004, 14184005, 14184006, 14184007, 14184008, 14184009, 14184010, 14184011, 14188009, 14188010, 14188012, 14188013, 14188014, 14188015, 14188016, 14188017, 14188018, 14191021, 14191022, 14191023, 14191024, 14191030, 14193002, 14193003, 14193007, 14193011, 14193012, 14193013, 14193014, 14193022, 14197042, 14197047, 14197048, 14197049. Expiry Datres Nov. 15, 2014 to Jan. 24, 2015.
510(k) numbers
not on file
PMA numbers
not on file
Product code
NPT
Firm FEI
3003882783
Firm city
Los Gatos
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register