Skip to the content

FDA Medical Device Recalls (openFDA)

EBI, L.P.: EBI Ball Tip Guide Wire 3.2mm x 55cm found in the Ankle Arthrodesis Nail Instr., Part Number 469055.

Recall number
Z-0916-05
Recalling firm
EBI, L.P.
Product
EBI Ball Tip Guide Wire 3.2mm x 55cm found in the Ankle Arthrodesis Nail Instr., Part Number 469055.
Status
Terminated
Initiated
2005-03-16
Posted
2005-06-22
Terminated
2006-06-30
Reason
The wire appears to be soft and the wire bent during insertion. This causes kinks in which the reamer could not pass over the bent/kinked wire.
Root cause
Other
Action
The firm distributed a recall letter to their EBI Fracture Sales Representatives and their EBI Trauma Distributors on 3/16/2005.
Quantity
280 units
Distribution
These devices are distributed nationwide to sales representatives who provide the product directly to the hospital/physician at the time of the surgery. The product was also distributed to Italy and Spain. There are no government accounts.
Lot, serial or model codes
Lot Number:   069811; 120055; 120056; 120057; 123827; 125637; 125638.
510(k) numbers
["K021786"]
PMA numbers
not on file
Product code
HXI
Firm FEI
2242816
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register