Skip to the content

FDA Medical Device Recalls (openFDA)

Kerr Corporation: Brand name: Maxcem Elite, Product: a dental cement, Part Number: 34418. The intended use of this device is for cementation of all indirect restorations including ceramic, resin and metal-based inlays, onlays, crowns, bridges, posts and veneers. Additional Indications include core-buildup material, pit and fissure sealant and cementation of crown restorations to implants

Recall number
Z-0915-2012
Recalling firm
Kerr Corporation
Product
Brand name: Maxcem Elite, Product: a dental cement, Part Number: 34418. The intended use of this device is for cementation of all indirect restorations including ceramic, resin and metal-based inlays, onlays, crowns, bridges, posts and veneers. Additional Indications include core-buildup material, pit and fissure sealant and cementation of crown restorations to implants
Status
Terminated
Initiated
2011-02-03
Posted
2012-01-27
Terminated
2012-01-31
Reason
The recall was initiated because Kerr Corporation has confirmed that the use of the affected lot of Maxcem Elite may not provide users with a minimum of 2 minutes of worktime thereby causing the product to prematurely set by 15 seconds.
Root cause
Nonconforming Material/Component
Action
Kerr Corporation sent a Urgent Medical Device Recall letter dated March 14, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. It has recently come to our attention that use of the affected lot of product may not provide users with a minimum of 2 minutes of worktime thereby causing the product to prematurely set by 15 seconds. If this were to happen prior to placing restorations on the tooth or core build-up the restorations may not adhere properly. If you have any of the affected product listed above please contact Kerr Customer Care at 1-800-537-7123 to receive an RMA number. The RMA will allow for a quick return and replacement or credit. Please fax back the enclosed Recall Return Form in order to confirm your receipt of this recall notification regardless of whether you have any product in your inventory.
Quantity
Four hundred and sixty-three (463)
Distribution
Worldwide Distribution -- USA (nationwide) and the countries of Canada, Australia and New Zealand.
Lot, serial or model codes
Lot Number 3360900
510(k) numbers
["K073209"]
PMA numbers
not on file
Product code
MZW
Firm FEI
2024312
Firm city
Orange
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register