Skip to the content

FDA Medical Device Recalls (openFDA)

Mako Surgical Corporation: Tibial Array and Femoral Array. Model Number(s): 0150ARR00001 and 0150ARR00002. Manufactured by: MAKO Surgical, Corp. The devices are accessories used with MAKO's Tactile Guidance System (TGS). The Tactile Guidance System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.

Recall number
Z-0915-2009
Recalling firm
Mako Surgical Corporation
Product
Tibial Array and Femoral Array. Model Number(s): 0150ARR00001 and 0150ARR00002. Manufactured by: MAKO Surgical, Corp. The devices are accessories used with MAKO's Tactile Guidance System (TGS). The Tactile Guidance System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
Status
Terminated
Initiated
2008-05-02
Posted
2009-02-11
Terminated
2009-08-21
Reason
The product in its current state may fail during use, which would cause the instrument to be in-accurate.
Root cause
Component design/selection
Action
A MAKO employee will visually verify product at each account for any product affected by this recall. In addition, a recall letter dated July 1, 2008 was mailed to users (surgeon and owner of system, e.g. hospital). Customers were instructed to complete the enclosed Acknowledgement form and return it to your Clinical Technical Specialist (CTS) or in the enclosed pre-paid addressed envelope to the Mako Surgical Corporation. Please direct questions to your CTS or call 954-927-2044, extension 604.
Quantity
44
Distribution
Nationwide Distribution USA including states of FL, NY, CA, TX, GA, TN, NC and PA.
Lot, serial or model codes
Femoral Tray - Model 0150ARR00001, Lot Number/Quantity: 0609038 (6), 06051107 (4), 01010806 (4), 09010108 (2), 06040707-01 (1), 01010107 (3), 01010607 (1), 01010106 (1) and 06040707 (2) and    Tibial Tray - Model 0150ARR00002, Lot Number/Quantity: 06060308 (6), 06061107 (7), 01020866 (2), 01010507 (1), 01020806 (1), 01020107 (3), 09081107 (2) and 01020607 (2).
510(k) numbers
["K072806"]
PMA numbers
not on file
Product code
HAW
Firm FEI
3005985723
Firm city
Davie
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register