Ortho-Clinical Diagnostics: enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges, Catalog # enGen -- Common/Usual Name: enGen(TM) Track System, Hettich Centrifuge -- Ortho-Clinical Diagnostics, Inc. Rochester, NY --- For in vitro diagnostic use only. The VITROS 5,1 FS Chemistry System with enGen(TM) Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).
- Recall number
- Z-0914-2012
- Recalling firm
- Ortho-Clinical Diagnostics
- Product
- enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges, Catalog # enGen -- Common/Usual Name: enGen(TM) Track System, Hettich Centrifuge -- Ortho-Clinical Diagnostics, Inc. Rochester, NY --- For in vitro diagnostic use only. The VITROS 5,1 FS Chemistry System with enGen(TM) Laboratory Automation System is intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).
- Status
- Terminated
- Initiated
- 2011-04-28
- Posted
- 2012-01-26
- Terminated
- 2018-07-02
- Reason
- Ortho Clinical Diagnostics received complaints of centrifuge buckets disconnecting from their supports during centrifugation when using the Hettich Rotanta 46 RSC Robotic Centrifuges, Models 4815, 4816 and 4817, that may be configured with enGen(TM) Laboratory Automation Systems.
- Root cause
- Component design/selection
- Action
- Ortho-Clinical Diagnostics sent an Urgent Product Correction Notification letter dated April 28, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to post the enclosed attachment near their centrifuge as a reference. Clean and carefully inspect the suspension groove area of the centrifuge buckets for cracks as part of their weekly maintenance procedure. Careful to check for micro-cracks that may not be easily detectable. If any cracks/micro-cracks, corrosion or damage are present, they must immediately discontinue using the centrifuge and contact OCD Customer Technical Services at 1-800-421-3311 to obtain replacements. OCD recommends that they do not exceed the manufacturer's relative centrifuge force (RCF) and run time limits for the blood collection devices used by their laboratory. Complete and return the attached Confirmation of Receipt Form. For questions regarding this recall call the Customer Service Center at 1-800-421-3311. .
- Quantity
- 78 units worldwide (18 Domestic and 68 Foreign)
- Distribution
- Worldwide Distribution - USA (nationwide) and the countries of Australia, Canada, Chile, England, France, Germany, Italy, Mexico, Singapore and Spain.
- Lot, serial or model codes
- All enGen(TM) Laboratory Automation Systems configured with Hettich Rotanta 46 RSC Robotic Centrifuges.
- 510(k) numbers
- K063144
- PMA numbers
- not on file
- Product code
- JJE
- Firm FEI
- 1000136573
- Firm city
- Rochester
- Firm state
- NY
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28