FDA Medical Device Recalls (openFDA)
Bard Peripheral Vascular Inc: Bard TruGuide Coxail Biopsy Needle with Depth Stop and additional Blunt Tip Stylet, Catalog Number C1816B, Device Listing Number E148687. The Bard TruGuide Coaxial Biopsy Needle is a three-part device consisting of an outer cannula with an attached female luer lock hub, and inner stylet with an attached male luer lock hub, and a flexible slip rink depth stop.
- Recall number
- Z-0913-06
- Recalling firm
- Bard Peripheral Vascular Inc
- Product
- Bard TruGuide Coxail Biopsy Needle with Depth Stop and additional Blunt Tip Stylet, Catalog Number C1816B, Device Listing Number E148687. The Bard TruGuide Coaxial Biopsy Needle is a three-part device consisting of an outer cannula with an attached female luer lock hub, and inner stylet with an attached male luer lock hub, and a flexible slip rink depth stop.
- Status
- Terminated
- Initiated
- 2006-05-01
- Posted
- 2006-05-31
- Terminated
- 2012-04-24
- Reason
- Manufacturing change that has introduced variability in the length of the inner stylet. When the inner stylet length is inserted into the cannula component, the exposed stylet length may be up to 8mm longer than expected.
- Root cause
- Other
- Action
- Recall letters were sent to all consignees (Radiology Department Manager with a copy to the Hospital Administrator) via Federal Express overnight delivery with signature receipt required on May 1, 2006. Each letter included a Recall and Effectiveness Check Form for each consignee to fill out and fax to Bard Peripheral Vascular, Inc (BPV).
- Quantity
- 225
- Distribution
- The product was distributed to the following five states: 1) California, 2) Florida 3) New Jersey, 3) Missouri, 4) Texas, 5) Washington. The product was also distributed to Canada and Latin America. Note: units distributed to Latin America remained in Bard''s distribution system and have been returned.
- Lot, serial or model codes
- Lot # REP JO196
- 510(k) numbers
- ["K915801","K936194"]
- PMA numbers
- not on file
- Product code
- FCG
- Firm FEI
- 2020394
- Firm city
- Tempe
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28