FDA Medical Device Recalls (openFDA)
Boston Scientific Corp: Boston Scientific EndoVive 24F (8 mm) Safety PEG Kit, Push Method for use with sterile Xylocaine ampule, latex free, sterile, Boston Scientific Corp., Natick, MA; Cat. no./REF 6649, inner pouch UPN M00566490 and outer carton UPN M00566491. The devices are indicated for providing nutrition directly into the stomach in adult and pediatric patients who are unable to consume nutrition by conventional means.
- Recall number
- Z-0912-2009
- Recalling firm
- Boston Scientific Corp
- Product
- Boston Scientific EndoVive 24F (8 mm) Safety PEG Kit, Push Method for use with sterile Xylocaine ampule, latex free, sterile, Boston Scientific Corp., Natick, MA; Cat. no./REF 6649, inner pouch UPN M00566490 and outer carton UPN M00566491. The devices are indicated for providing nutrition directly into the stomach in adult and pediatric patients who are unable to consume nutrition by conventional means.
- Status
- Terminated
- Initiated
- 2008-12-30
- Posted
- 2009-01-28
- Terminated
- 2009-10-30
- Reason
- The user may be unable to pass the device over the guidewire for placement in the patient, resulting in an inability to use the device.
- Root cause
- Other
- Action
- Consignees were notified by "Urgent Medical Device Recall" letter dated 12/30/08. The letter described the problem with the product, instructed customers to determine whether they store these products in the outer carton or the inner pouch at their facility, to locate all affected products, and to return them to the firm. For additional information, contact Boston Scientific Corporation at 1-508-683-4678.
- Quantity
- 2460 of all products
- Distribution
- Nationwide.
- Lot, serial or model codes
- Inner pouch (UPN M00566490) lots 12007899 and 12101506 packaged in outer cartons (UPN M00566491) with lot numbers of 12065904 and 12148450.
- 510(k) numbers
- ["K031538"]
- PMA numbers
- not on file
- Product code
- KNT
- Firm FEI
- 1828132
- Firm city
- Spencer
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28