Skip to the content

FDA Medical Device Recalls (openFDA)

ADAC Laboratories: Pinnacle3 Radiation Therapy Planning System, Version 6.6c Kit, Model Number 9200-0613C-ENG

Recall number
Z-0912-04
Recalling firm
ADAC Laboratories
Product
Pinnacle3 Radiation Therapy Planning System, Version 6.6c Kit, Model Number 9200-0613C-ENG
Status
Terminated
Initiated
2004-03-03
Posted
2004-07-20
Terminated
2005-02-01
Reason
A software defect, relating to the motorized wedge functionality when using an Elekta linear accelerator, can result in incorrect dose delivery to the patient. Also, dose delivery in the wrong location may occur when using a Elekta Leksell headframe.
Root cause
Other
Action
A notification dated March 3, 2004 was sent to the customers. The notification gave recommendations for avoiding the problem and stated that updated software would be provided when it becomes available. Firm representatives made follow-up telephone calls to the consignees.
Quantity
55
Distribution
The recalled products were shipped to consignees located nationwide the United States and in Australia, Canada, China, and Korea.
Lot, serial or model codes
Version 6.6c
510(k) numbers
not on file
PMA numbers
not on file
Product code
MUJ
Firm FEI
3004022368
Firm city
Madison
Firm state
WI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register