FDA Medical Device Recalls (openFDA)
Medtronic Inc. Cardiac Rhythm Managment: Medtronic Paceart 2004 Second Edition & Paceart 2005 First Edition versions of the Paceart Gateway Software. The Medtronic Paceart System is a personal computer based pacemaker-testing system, a 12-lead electrocardiograph, and a transtelephonic receiving station. It provides data that can be used to analyze implanted pacemaker performance based on electrocardiographic measurements, either taken directly from the patient or programmer. The system can measure, store, and display any of the 12 standard leads. Reports and charts are available by means of a laser printer. The system also includes a database that collects and stores patient data. Optional software allows the device to function as an unattended transtelephonic receiving station.
- Recall number
- Z-0911-06
- Recalling firm
- Medtronic Inc. Cardiac Rhythm Managment
- Product
- Medtronic Paceart 2004 Second Edition & Paceart 2005 First Edition versions of the Paceart Gateway Software. The Medtronic Paceart System is a personal computer based pacemaker-testing system, a 12-lead electrocardiograph, and a transtelephonic receiving station. It provides data that can be used to analyze implanted pacemaker performance based on electrocardiographic measurements, either taken directly from the patient or programmer. The system can measure, store, and display any of the 12 standard leads. Reports and charts are available by means of a laser printer. The system also includes a database that collects and stores patient data. Optional software allows the device to function as an unattended transtelephonic receiving station.
- Status
- Terminated
- Initiated
- 2006-02-02
- Posted
- 2006-05-31
- Terminated
- 2006-12-31
- Reason
- The Paceart System could inappropriately insert data from a patient's Medtronic CareLink transmission into another patient's Paceart record. An issue associated with the automated batch transfer of implanted cardiac device data from the Medtronic CareLink network to the Paceart System exists.
- Root cause
- Other
- Action
- Affected user and potential users of the Gateway product were notified via phone and/ or a customer notification. The notification offers interim options until a software update is available. Customers are asked to respond back to Metronic with an enclosed Acknowledgement form.
- Quantity
- 210
- Distribution
- Nationwide
- Lot, serial or model codes
- Paceart 2004 second edition and Paceart 2005 first edition
- 510(k) numbers
- ["K024278"]
- PMA numbers
- not on file
- Product code
- DPS
- Firm FEI
- 2182208
- Firm city
- Minneapolis
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28