Skip to the content

FDA Medical Device Recalls (openFDA)

Nipro Diagnostics, Inc.: MooreMedical Supporting Health & Care...MooreBrand Blood Glucose Monitor for Self-Testing & Multi-Patient Use TRUEtrack. Manufactured by: NIPRO DIAGNOSTICS, Fort Lauderdale, FL 33309 U.S.A. Is intended for the quantitative determination of glucose in human whole blood taken from finger or forearm.

Recall number
Z-0910-2014
Recalling firm
Nipro Diagnostics, Inc.
Product
MooreMedical Supporting Health & Care...MooreBrand Blood Glucose Monitor for Self-Testing & Multi-Patient Use TRUEtrack. Manufactured by: NIPRO DIAGNOSTICS, Fort Lauderdale, FL 33309 U.S.A. Is intended for the quantitative determination of glucose in human whole blood taken from finger or forearm.
Status
Terminated
Initiated
2013-10-02
Posted
2014-02-03
Terminated
2016-01-28
Reason
Manufactured with an incorrect factory set unit of measure.
Root cause
Software design (manufacturing process)
Action
Nipro Diagnostics sent an Urgent Medical Device Voluntary Recall letter dated December 31, 2013 to all affected customers. The customers were instructed to do the following: 1) Screen their inventory immediately for any of the serial numbers listed on the back of the products. 2) If they have any of the affected products, return the product by contacting Stericycle at 1-866-236-4518. 3) Complete and return the enclosed response form via fax at 1-888-358-7294. Customers with questions were instructed to call Stericycle at 1-866-236-4518. For questions regarding this recall call 954-677-9201.
Quantity
105
Distribution
Worldwide Distribution - USA (nationwide, HI and Puerto Rico and Internationally to Canada, European Union, Jamaica, Suriname, Trinidad and Tobago, Kazakhastan, Russian, Turkey, Australia, Bangladesh, Pakistan, Cambodia, Laos, Myanmar, Thailand, Vietnam, and India.
Lot, serial or model codes
Finished Goods: KJ0731; Lot Numbers: KK0964, KL01213, KL01233-3TI, KL0267-10, KL0381-14, KL0637-2TI, KL0716TI, KL0860TI, KP0074TI.
510(k) numbers
["K032657"]
PMA numbers
not on file
Product code
NBW
Firm FEI
1000113657
Firm city
Ft Lauderdale
Firm state
FL
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register