Skip to the content

FDA Medical Device Recalls (openFDA)

Ebi, Llc: Cypher MIS Screw System-The system includes screws, various types and sizes of rods, and set screws. Various instruments are also available for use by the surgeon to facilitate implantation of the device. Non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion.

Recall number
Z-0909-2015
Recalling firm
Ebi, Llc
Product
Cypher MIS Screw System-The system includes screws, various types and sizes of rods, and set screws. Various instruments are also available for use by the surgeon to facilitate implantation of the device. Non-cervical spinal fixation device intended for immobilization and stabilization as an adjunct to fusion.
Status
Terminated
Initiated
2014-10-08
Posted
2015-01-05
Terminated
2015-12-21
Reason
Inoperative screw head seat splay and/or fracture as a result of positioning of the rod which is a part of the Cypher MIS Screw System.
Root cause
Device Design
Action
The firm, EBI LLC/Biomet, sent an "URGENT PRODUCT INFORMATION" letter dated 10/6/2014 via Fed Ex to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to follow the steps outlined in the letter. If you have any questions, please feel free to contact Director of Regulatory Affairs at 303-501-8548.
Quantity
2816 approx.units ( 2688 units US and 128 OUS)
Distribution
Worldwide Distribution: US Distribution to states of: AL, CA, CT, IN, LA, NC, NJ TX & WI; and the country of: The Netherlands.
Lot, serial or model codes
Item Number: 14-571845 5.5 TI 10.5X45MM CANN TRANS 14-571850 5.5 TI 10.5X50MM CANN TRANS 14-571855 5.5 TI 10.5X55MM CANN TRANS 14-571860 5.5 TI 10.5X60MM CANN TRANS 14-571865 5.5 TI 10.5X65MM CANN TRANS 14-571870 5.5 TI 10.5X70MM CANN TRANS 14-571875 5.5 TI 10.5X75MM CANN TRANS 14-571880 5.5 TI 10.5X80MM CANN TRANS 14-571885 5.5 TI 10.5X85MM CANN TRANS 14-571890 5.5 TI 10.5X90MM CANN TRANS 14-571895 5.5 TI 10.5X95MM CANN TRANS 14-571899 5.5 TI 10.5X100MM CANN TRANS
510(k) numbers
["K123549"]
PMA numbers
not on file
Product code
NKB
Firm FEI
2242816
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register