FDA Medical Device Recalls (openFDA)
Bard Access Systems: Hickman 9.0 French Dual Lumen Catheter with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer system (pediatric product code 0600580; adult product code 0600600). Product Usage: Hickman and Leonard Catheters are designed for long-term vascular access and for use in patients that lack adequate peripheral venous access.
- Recall number
- Z-0909-2013
- Recalling firm
- Bard Access Systems
- Product
- Hickman 9.0 French Dual Lumen Catheter with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer system (pediatric product code 0600580; adult product code 0600600). Product Usage: Hickman and Leonard Catheters are designed for long-term vascular access and for use in patients that lack adequate peripheral venous access.
- Status
- Terminated
- Initiated
- 2013-02-21
- Posted
- 2013-03-06
- Terminated
- 2014-01-03
- Reason
- Bard Access Systems is recalling two lots of Hickman 9.0 French and one lot of Leonard 10.0 French Dual Lumen Catheters with SureCuff Tissue Ingrowth Cuff with Peel-Apart Percutaneous Introducer Systems due to the wrong size introducer being included with the kit.
- Root cause
- Mixed-up of materials/components
- Action
- Bard sent an Urgent Recall Notification letters dated February 2013 to all affected customers. The letter identified the affected product, problem, and actions to be taken. Customers were instructed to examine their inventory for the affected product, return any remaining in stock and complete the enclosed Reply Form and Inventory Reconciliation Form. For questions contact Bard Access Systems Customer Service at 1-800-290-1689.
- Quantity
- 195 units.
- Distribution
- USA Nationwide Distribution and the countries Europe, Belgium, Australia, and Latin America.
- Lot, serial or model codes
- Lot numbers: HUWI1379 and HUWI1380.
- 510(k) numbers
- ["K830233"]
- PMA numbers
- not on file
- Product code
- DYB
- Firm FEI
- 3006260740
- Firm city
- Salt Lake City
- Firm state
- UT
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28