FDA Medical Device Recalls (openFDA)
Medtronic Cardiovascular Surgery-the Heart Valve Division: Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic Porcine Aortic Bioprostheses.
- Recall number
- Z-0908-2014
- Recalling firm
- Medtronic Cardiovascular Surgery-the Heart Valve Division
- Product
- Product Name: Model 7305- Mosaic Aortic Obturator The Mosaic Aortic Obturators are used with the Mosaic Porcine Aortic Bioprostheses.
- Status
- Terminated
- Initiated
- 2014-01-09
- Posted
- 2014-02-03
- Terminated
- 2014-05-30
- Reason
- Medtronic, Inc. initiated a recall of Medtronic Mosaic Aortic Obturator Model 7305 used with the Mosaic Porcine Aortic Bioprosthesis, Model Number 305, because of higher-than-expected transvalvular gradients occurring post implant.
- Root cause
- Device Design
- Action
- Medtronic sent an Customer Notification letter dated January 9, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customers. Medtronic recommends surgeons continue to remain diligent in selecting the appropriate size of Mosaic Aortic Bioprosthesis. Adoption of the updated Mosaic Aortic Bioprosthesis sizing chart (iEOA chart) and Mosaic Obturators/Sizers will facilitate optimal aortic valve size selection. Your Medtronic Field Representative will be contacting you about receiving a new Mosaic Aortic Obturator/Sizer set and sizing chart. If you would prefer to order the new Mosaic Aortic Obturator/Sizer set directly through Medtronic Customer Service, contact them at 1-800-848-9300. Medtronic will notify all applicable regulatory agencies, as required, about this matter. Please share this notification with others in your organization as appropriate. We appreciate your review of this notification and apologize for the inconvenience that it may cause. If you have any questions, please contact your Medtronic Field Representative or Lifeline Technical Services at 1-877-526-7890.
- Quantity
- 8276 units
- Distribution
- Worldwide Distribution - USA (nationwide)
- Lot, serial or model codes
- Model 7305. Lot/Serial Numbers: The catalog numbers for the Obturator/Sizer are 7305C, 7305UX, and 7305OD. Scope of this notification includes all Mosaic Obturators/Sizers, used with Mosaic Porcine Aortic Bioprosthesis implants.
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- LWR
- Firm FEI
- 2025587
- Firm city
- Santa Ana
- Firm state
- CA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28