Skip to the content

FDA Medical Device Recalls (openFDA)

Bio-Rad Laboratories: Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Control, 1 Low Positive Control, 1 Substrate, 1 Stop Reagent. Labeled: Distributed by: Bio-Rad Laboratories-Redmond, WA 98052 For the qualitative, semi-quantitative and quantitative detection of human IgG antibodies to rubella virus in human serum by enzyme immunoassay, to aid in the assessment to the patient's immunological response to rubella, and as a qualitative screening test to determine the immune status of individuals, including women of childbearing age.

Recall number
Z-0908-2010
Recalling firm
Bio-Rad Laboratories
Product
Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Control, 1 Low Positive Control, 1 Substrate, 1 Stop Reagent. Labeled: Distributed by: Bio-Rad Laboratories-Redmond, WA 98052 For the qualitative, semi-quantitative and quantitative detection of human IgG antibodies to rubella virus in human serum by enzyme immunoassay, to aid in the assessment to the patient's immunological response to rubella, and as a qualitative screening test to determine the immune status of individuals, including women of childbearing age.
Status
Terminated
Initiated
2008-08-20
Posted
2010-03-02
Terminated
2010-04-08
Reason
Elevated optical density (OD) values with the kit Negative Control.
Root cause
Employee error
Action
An "Urgent Product Correction" letter dated August 26, 2008 was sent via overnight carrier to all customers. The letter described the affected product, problem, and action to be taken by customer. The customers are to review their current inventory of the product for any remaining kits of the specified lot number. The customer is to complete the form and fax to the Technical Support department if replacement kits were needed. The remaining kits will be replaced once the form is completed and faxed. If you have any questions, please contact Technical Support at 1-800-224-6723, options 2 then 4.
Quantity
82 kits
Distribution
Nationwide distribution: NE, NY, FL, GA, MA, OR, WI, MN, MI, MS, NC, CA, MD, MO, IA, DC
Lot, serial or model codes
Lot F1108
510(k) numbers
["K961053"]
PMA numbers
not on file
Product code
LFX
Firm FEI
3008272466
Firm city
Redmond
Firm state
WA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register