FDA Medical Device Recalls (openFDA)
Bio-Rad Laboratories: Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Control, 1 Low Positive Control, 1 Substrate, 1 Stop Reagent. Labeled: Distributed by: Bio-Rad Laboratories-Redmond, WA 98052 For the qualitative, semi-quantitative and quantitative detection of human IgG antibodies to rubella virus in human serum by enzyme immunoassay, to aid in the assessment to the patient's immunological response to rubella, and as a qualitative screening test to determine the immune status of individuals, including women of childbearing age.
- Recall number
- Z-0908-2010
- Recalling firm
- Bio-Rad Laboratories
- Product
- Rubella IgG EIA test kit, containing 1 Rubella IgG Microplate, 1 Negative Control, 1 High Positive Control, 1 Low Positive Control, 1 Substrate, 1 Stop Reagent. Labeled: Distributed by: Bio-Rad Laboratories-Redmond, WA 98052 For the qualitative, semi-quantitative and quantitative detection of human IgG antibodies to rubella virus in human serum by enzyme immunoassay, to aid in the assessment to the patient's immunological response to rubella, and as a qualitative screening test to determine the immune status of individuals, including women of childbearing age.
- Status
- Terminated
- Initiated
- 2008-08-20
- Posted
- 2010-03-02
- Terminated
- 2010-04-08
- Reason
- Elevated optical density (OD) values with the kit Negative Control.
- Root cause
- Employee error
- Action
- An "Urgent Product Correction" letter dated August 26, 2008 was sent via overnight carrier to all customers. The letter described the affected product, problem, and action to be taken by customer. The customers are to review their current inventory of the product for any remaining kits of the specified lot number. The customer is to complete the form and fax to the Technical Support department if replacement kits were needed. The remaining kits will be replaced once the form is completed and faxed. If you have any questions, please contact Technical Support at 1-800-224-6723, options 2 then 4.
- Quantity
- 82 kits
- Distribution
- Nationwide distribution: NE, NY, FL, GA, MA, OR, WI, MN, MI, MS, NC, CA, MD, MO, IA, DC
- Lot, serial or model codes
- Lot F1108
- 510(k) numbers
- ["K961053"]
- PMA numbers
- not on file
- Product code
- LFX
- Firm FEI
- 3008272466
- Firm city
- Redmond
- Firm state
- WA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28