FDA Medical Device Recalls (openFDA)

Getinge/Castle, Inc.: Model 8666 Washer/Disinfector with holding tank heater option. Firm on label is Getinge Disinfection AB, Vaxjo, Sweden.

Recall number
Z-0908-03
Recalling firm
Getinge/Castle, Inc.
Product
Model 8666 Washer/Disinfector with holding tank heater option. Firm on label is Getinge Disinfection AB, Vaxjo, Sweden.
Status
Terminated
Initiated
2003-03-25
Posted
2003-06-05
Terminated
2004-01-14
Reason
Hot water may splash from the washer chamber creating a burn hazard.
Root cause
Device Design
Action
Letter dated 3/25/03 flagged: ''Important Safety Alert/Product Upgrade''. Customers advised to discontinue use until the field correction is performed.
Quantity
75 units
Distribution
Nationwide. Medical facilities in AL, CA, CO, LA, MA, MI, MN, MO, NC, NH, NM, OK, PA, TX, WA, and WI.
Lot, serial or model codes
Serial numbers 81972 through 82063; and SEV0141098 through SEV0304112.
510(k) numbers
["K920413"]
PMA numbers
not on file
Product code
MEC
Firm FEI
1314329
Firm city
Rochester
Firm state
NY
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register