Ventana Medical Systems, Inc.: VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001
- Recall number
- Z-0907-2023
- Recalling firm
- Ventana Medical Systems, Inc.
- Product
- VENTANA PD-L1 (SP142) VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 07709374001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 08008540001 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 07011571001
- Status
- Open, Classified
- Initiated
- 2022-11-18
- Posted
- 2023-01-06
- Terminated
- not on file
- Reason
- Internal comparison studies revealed unacceptable, light staining. Light staining affects the borderline of positive versus negative test results. It can cause a reduction in both staining intensity and the percent positivity, which could potentially lead to a false-negative PD-L1 result.
- Root cause
- Process design
- Action
- On 11/18/2022, the firm sent an "URGENT MEDICAL DEVICE CORRECTION", Letter via UPS ground to customer to inform them that internal comparison studies revealed unacceptable, light staining with some VENTANA PD-L1 (SP142) lots. This can cause a reduction in both staining intensity and percent positivity, which could potentially lead to a false-negative PD-L1 result. Actions required of Customers: 1) Immediately discontinue the use of the VENTANA PD-L1 (SP142) lot numbers listed in Table 2 and discard any remaining inventory according to your local guidelines. 2) Be aware of the new expiration dates for the VENTANA PD-L1 (SP142) lot numbers listed in Tables 3 and 4. 3) Review negative results for the lots listed in Tables 2 and 3 as directed. 4) Follow standard laboratory operating procedures to investigate any suspected false negative results generated with the affected lots. 5) Consult with the physician or pathologist at their facility to determine specific clinical implications for their patients. 6) Refer to the Product Credit section on the last page of this UMDC for the type of credit available to them and their actions needed. 7) Complete all sections of the enclosed Recall Response Form (TP-01699) and email it to: roche7953@sedgwick.com even if they are not requesting product credit. 8) Put PD-L1 (SP142) in the subject line of the email. Or fax it to 1-888-345-0279. 9) Be aware product will not be sent nor will credit be processed until the Recall Response Form is received. 10) Post this UMDC on or near the affected products/lots as a reminder to continuously check the expiration date and the instruction. 11) File this Urgent Medical Device Correction (UMDC) for future reference. For questions, contact the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-227-2155.
- Quantity
- 8,927 kits
- Distribution
- Worldwide distribution - U.S. Nationwide distribution including in the states of AL, AZ, CA, CO,CT, DC, DE, FL, GA, IL, IN, KS, KY, MA, MD, MI, MN, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, and WI. The countries of Canada, Austria, Belgium, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Ireland, Italy, Latvia, Lithuania, Luxembourg, Netherlands, Norway, Poland, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom, Albania, Armenia, Bangladesh, Belarus, Costa Rica, Cote d'Ivoire, Cuba, Dominican Republic, Egypt, El Salvador, Ghana, Guatemala, Iran, Israel, Jordan, Kenya, Kosovo, Kuwait, Lebanon, Mauritius, Morocco, Myanmar, Nigeria, North Macedonia, Oman, Panama, Philippines, Russian Fed., Saudi Arabia, Serbia, South Africa, Tanzania, Turkey, Turkmenistan and United Arab Emirates. Update (04-Jan-2023): Argentina, Australia, Brazil, Chile, China, Colombia, Ecuador, Hong Kong, Honduras, Indonesia, India, Japan, Korea, Mexico, Malaysia, New Zealand, Peru, Pakistan, Paraguay, Singapore, Thailand, Taiwan, Uruguay and Vietnam.
- Lot, serial or model codes
- UDI-DI Code: 04015630967155 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 F30596 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G27114 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 G31438 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H03958 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H13049 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 H35110 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J12769 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J18775 VENTANA PD-L1 (SP142) Assay (Class III, US IVD) 7709374001 J25361 UDI-DI Code: 07613336106709 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 F31427 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G06715 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G16831 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 G24564 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H00001 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H07164 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H15132 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H18578 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 H23050 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J00572 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J01568 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J14244 VENTANA PD-L1 (SP142) Assay (CE IVD Predictive) 8008540001 J20989 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H08373 UDI-DI Code: 04015630976591 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H17865 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H30521 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 H35111 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 J11932 VENTANA PD-L1 (SP142) IHC Assay (CE IVD Analytical) 7011571001 J20132
- 510(k) numbers
- not on file
- PMA numbers
- ["P160002S017"]
- Product code
- PLS
- Firm FEI
- 1000125183
- Firm city
- Oro Valley
- Firm state
- AZ
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28