FDA Medical Device Recalls (openFDA)

Guidant Corporation: CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds). Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798. USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.

Recall number
Z-0907-05
Recalling firm
Guidant Corporation
Product
CONTAK RENEWAL (model H135) Heart Failure Device Cardiac Resynchronization Therapy Defibrillators (CRT-Ds). Sterilized with gaseous ethylene oxide. Guidant Corporation, Cardiac Rhythm Management, 4100 Hamline Avenue North, St. Paul, MN 55112-5798. USA. Renewal provides ventricular tachyarrhythmia and cardiac resynchronization therapies.
Status
Terminated
Initiated
2005-06-17
Posted
2005-06-30
Terminated
2007-01-10
Reason
Laboratory analysis revealed that a deterioration in a wire insulator within the lead connector block, in conjunction with other factors, could cause a short circuit and loss of device function due to diversion of therapy energy away from the heart and into device circuitry.
Root cause
Other
Action
Dear Doctor letter was dated and issued 06/17/05 informing Doctors of an shorting issue of the Renewal 1 & 2 CRT-D devices. The letter gave indications of device failure and recommendations for patient visits. A press release was also issued on 06/17/05.
Quantity
16,000 (Approximately 11,900 remain implanted worldwide)
Distribution
worldwide. Including the United States and countries such as: Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, Colombia, Cyprus, Czech Republic, Czechoslovakia, Denmark, Finland, France, Germany, Hong Kong, India, Ireland, Israel, Italy, Jordan, Kuwait, Lebanon, Luxembourg, Malaysia, Mexico, Monaco, Netherlands, New Zealand, Norway, Portugal, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Syrian Arab Republic, Thailand, Turkey, United Kingdom and Venezuela.
Lot, serial or model codes
RENEWAL (model H135) serial number range 100027 - 999983 includes CRT-Ds manufactured on or before August 26, 2004. These ranges include all affected devices but may include some unaffected devices.
510(k) numbers
not on file
PMA numbers
["P010012S002"]
Product code
NIK
Firm FEI
2124215
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register