FDA Medical Device Recalls (openFDA)
Olympus Corporation of the Americas: Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.
- Recall number
- Z-0905-2025
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Olympus MAJ-891 Forceps/Irrigation Plug (Isolated Type). Model/Catalog Number: MAJ-891. This instrument is designed to be attached to the instrument channel port of Olympus endoscopes to allow both irrigation and the use of endo-therapy accessories.
- Status
- Open, Classified
- Initiated
- 2024-12-18
- Posted
- 2025-01-16
- Terminated
- not on file
- Reason
- Potential for inadvertent retention of biomaterial when the reprocessing procedure is not performed as instructed, so as not completely removing and disassembling the device.
- Root cause
- Labeling design
- Action
- Olympus notified consignees on about 12/18/2024 via letter. Consignees were instructed to utilize alternative devices to the MAJ-891 according to the provided compatibility table provided, ensure all personnel are completely knowledgeable and thoroughly trained on the content of this letter and the MAJ-891 reprocessing instructions, acknowledge receipt of the letter through the Olympus portal, https://olympusamerica.com/recall, and notify customers if the units have been further distributed. Additionally, Olympus will contact customers to schedule a mutually beneficial time for an Olympus Field Representative to visit each facility and review the contents of this letter and the MAJ-891 reprocessing instructions.
- Quantity
- 30,140 units
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Model Number: MAJ-891. UDI-DI: 04953170063114. All lot numbers.
- 510(k) numbers
- ["K912120"]
- PMA numbers
- not on file
- Product code
- FBN
- Firm FEI
- 2429304
- Firm city
- Center Valley
- Firm state
- PA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as