Skip to the content

FDA Medical Device Recalls (openFDA)

Randox Laboratories Ltd.: Liquid Assayed Specific Protein Control Level 3- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2684

Recall number
Z-0905-2023
Recalling firm
Randox Laboratories Ltd.
Product
Liquid Assayed Specific Protein Control Level 3- In vitro diagnostic use, in the quality control of serum on clinical chemistry and immunoassay systems Catalog Number: PS2684
Status
Open, Classified
Initiated
2022-11-22
Posted
2023-01-05
Terminated
not on file
Reason
Concentration of Rheumatoid Factor has decreased in Specific Protein Control Level 1-3 causing delay in reporting results due to Quality Controls running low outside range
Root cause
Under Investigation by firm
Action
Randox Laboratories Ltd issued via email to the US Distributor Randox Laboratories-US, Ltd., Kearneysville, West Virginia on 11/22/22. The US Distributor contacted the customers by letter on 11/28/22. The Urgent Medical Device Correction letter states reason for recall, health risk and action to take: Discard previous IFUs and download the updated IFUs from randox.com Review results generated with the affected batches in line with the clinical profile of the patient. Discuss the contents of this notice with your Medical Director. Complete and return the response form 12187-QA to technical.services@randox.com within five working days. Transmission of Field Safety Notice: Send a copy of the FSN to all affected customers and to those who need to be aware within your organization.
Quantity
49 units US
Distribution
US Nationwide distribution in the states of CA, FL, OH,ME, MO, NC, NJ, PA, TN.
Lot, serial or model codes
GTIN: 05055273204919 Lot Number/Exp. Date: 586LPC 28th Jun 23; 602LPC 28th Nov 23; 628LPC 28th Dec 24;
510(k) numbers
["K972608"]
PMA numbers
not on file
Product code
JJY
Firm FEI
1000361607
Firm city
Crumlin
Firm state
North
Firm country
Ireland

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register