Zimmer Inc.: Zimmer TIV Self Tap Screw, 42 mm. Manufactured by Zimmer Manufacturing B.V., A Subsidiary of Zimmer, Inc., Warsaw, IN; Item No. 48231904200. NOTE: This product is not sold in the United States. The device is used to attach healing plates to long bones for stabilization of fractures during the normal healing process. Devices are individually packaged and sold sterile.

Recall number
Z-0905-2009
Recalling firm
Zimmer Inc.
Product
Zimmer TIV Self Tap Screw, 42 mm. Manufactured by Zimmer Manufacturing B.V., A Subsidiary of Zimmer, Inc., Warsaw, IN; Item No. 48231904200. NOTE: This product is not sold in the United States. The device is used to attach healing plates to long bones for stabilization of fractures during the normal healing process. Devices are individually packaged and sold sterile.
Status
Terminated
Initiated
2008-12-12
Posted
2009-01-26
Terminated
2009-10-30
Reason
The package may contain a 40 mm long screw instead of the labeled 42 mm long screw, or vice versa.
Root cause
Process control
Action
The firm's facility in Japan was notified via letter, "Urgent: Device Recall" (dated 12/11/08) and requested to complete an enclosed Inventory Return Certification form. The letter instructed distributors to notify accounts to the user level to stop using the affected implants and return them to the Zimmer Distribution Center in the U.S. and fax (574-372-4265) a copy of the completed Inventory Return Certification form to Zimmer, Inc. For questions about the recall, please contact Zimmer, Inc. at 1-800-613-6131.
Quantity
408
Distribution
International Distribution including country of Japan.
Lot, serial or model codes
Lots 60548987, 60548986, 60548999, 60557220 and 60557219.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HWC
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register