ev3 Inc: Medtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090, Size: 5MM X 40MM X 90CM; 4) AB14W050040150, Size: 5MM X 40MM X 150CM; 5) AB14W050060150, Size: 5MM X 60MM X 150CM; 6) AB14W050080090, Size: 5MM X 80MM X 90CM; 7) AB14W050080150, Size: 5MM X 80MM X 150CM; 8) AB14W050100150, Size: 5MM X 100MM X 150CM; 9) AB14W050120150, Size: 5MM X 120MM X 150CM; 10) AB14W050150150, Size: 5MM X 150MM X 150CM; 11) AB14W050200090, Size: 5MM X 200MM X 90CM; 12) AB14W050200150, Size: 5MM X 200MM X 150CM; 13) AB14W060040090, Size: 6MM X 40MM X 90CM; 14) AB14W060040150, Size: 6MM X 40MM X 150CM; 15) AB14W060060150, Size: 6MM X 60MM X 150CM; 16) AB14W060080150, Size: 6MM X 80MM X 150CM; 17) AB14W060100150, Size: 6MM X 100MM X 150CM; 18) AB14W060120150, Size: 6MM X 120MM X 150CM; 19) AB14W060150150, Size: 6MM X 150MM X 150CM; 20) AB14W060200150, Size: 6MM X 200MM X 150CM; Balloon Dilatation Catheter

Recall number
Z-0904-2025
Recalling firm
ev3 Inc
Product
Medtronic NanoCross Elite PTA catheter, REF numbers amd sizes: 1) AB14W050020090, Size: 5MM X 20MM X 90CM; 2) AB14W050020150, Size: 5MM X 20MM X 150CM; 3) AB14W050040090, Size: 5MM X 40MM X 90CM; 4) AB14W050040150, Size: 5MM X 40MM X 150CM; 5) AB14W050060150, Size: 5MM X 60MM X 150CM; 6) AB14W050080090, Size: 5MM X 80MM X 90CM; 7) AB14W050080150, Size: 5MM X 80MM X 150CM; 8) AB14W050100150, Size: 5MM X 100MM X 150CM; 9) AB14W050120150, Size: 5MM X 120MM X 150CM; 10) AB14W050150150, Size: 5MM X 150MM X 150CM; 11) AB14W050200090, Size: 5MM X 200MM X 90CM; 12) AB14W050200150, Size: 5MM X 200MM X 150CM; 13) AB14W060040090, Size: 6MM X 40MM X 90CM; 14) AB14W060040150, Size: 6MM X 40MM X 150CM; 15) AB14W060060150, Size: 6MM X 60MM X 150CM; 16) AB14W060080150, Size: 6MM X 80MM X 150CM; 17) AB14W060100150, Size: 6MM X 100MM X 150CM; 18) AB14W060120150, Size: 6MM X 120MM X 150CM; 19) AB14W060150150, Size: 6MM X 150MM X 150CM; 20) AB14W060200150, Size: 6MM X 200MM X 150CM; Balloon Dilatation Catheter
Status
Open, Classified
Initiated
2024-12-10
Posted
2025-01-15
Terminated
not on file
Reason
There is the potential for in-vivo balloon deflation issues, including difficulty or inability to deflate the balloon in specific lots.
Root cause
Process change control
Action
Medtronic issued an URGENT: MEDICAL DEVICE RECALL notice to its consignees on 12/10/2024 via mail courier. The notice explained the issue, potential harms, and requested the following actions be taken: "" Immediately locate and quarantine all affected and unused NanoCross devices as listed in the provided scope. " Return all unused affected products in your inventory to Medtronic. Contact Medtronic Vascular Customer Service at 800-716-6700 to initiate a product return. Your local Medtronic Representative can assist you with the initiation of the return. " Complete and return the Customer Confirmation Form enclosed with this letter, acknowledging that you have received this information, and return it to RS.CFQFCA@medtronic.com within 30-days of receipt of this letter. " Please forward this notice to all those who need to be aware within your organization and maintain a copy for your records."
Quantity
5189 units
Distribution
Worldwide distribution - US (& PR) Nationwide and the countries of Chile, Cyprus, Estonia, Germany, Italy, Japan, Jordan, Puerto Rico, Spain, United States, Uzbekistan, and Yemen.
Lot, serial or model codes
1) AB14W050020090, UDI/DI 00763000758035, Serial Numbers: B738596, B749440, B764968; 2) AB14W050020150, UDI/DI 00763000758042, Serial Numbers: B761521, B781264, B794814; 3) AB14W050040090, UDI/DI 00763000758059, Serial Numbers: B740725, B748628; 3) AB14W050040090, UDI/DI 04987742337891, Serial Numbers: B740725; 4) AB14W050040150, UDI/DI 00763000758066, Serial Numbers: B748644, B755679, B757663, B760286, B761769, B762343, B775924, B775972, B776408, B792995; 4) AB14W050040150, UDI/DI 04987742337907, Serial Numbers: B755679; 5) AB14W050060150, UDI/DI 00763000758080, Serial Numbers: B755683, B761779, B762646, B792990; 6) AB14W050080090, UDI/DI 00763000758097, Serial Numbers: B797860; 7) AB14W050080150, UDI/DI 00763000758103, Serial Numbers: B734091, B750081, B755684, B761172, B761754, B762256, B775959, B793874; 8) AB14W050100150, UDI/DI none, Serial Numbers: B750054, B758860, B761368, B764977, B765527, B769645, B781316, B791878, B805118; 9) AB14W050120150, UDI/DI 00763000758141, Serial Numbers: B731160, B761782, B762036, B768737, B793229; 10) AB14W050150150, UDI/DI 00763000758165, Serial Numbers: B727462, B736200, B750625, B758294, B758866, B760303, B762032, B763323, B763333, B763355, B763360, B776072, B792620, B793222; 11) AB14W050200090, UDI/DI 00763000758172, Serial Numbers: B760301, B789934; 11) AB14W050200090, UDI/DI 04987742337815, Serial Numbers: B760301; 12) AB14W050200150, UDI/DI 00763000758189, Serial Numbers: B734116, B740732, B749498, B749509, B754865, B754876, B754889, B754921, B754938, B754944, B762313, B762346, B762623, B762814, B762967, B781392, B791857, B792372, B798527; 13) AB14W060040090, UDI/DI 00763000758219, Serial Numbers: B730934, B795892; 14) AB14W060040150, UDI/DI 00763000758226, Serial Numbers: B737344, B748659, B755681, B760297, B761167, B761194, B764136, B764985, B765532, B791881, B793013; 15) AB14W060060150, UDI/DI 00763000758240, Serial Numbers: B762430, B762853, B791832; 16) AB14W060080150, UDI/DI 00763000758264, Serial Numbers: B755686, B761181, B762366, B762433, B762921, B793869; 17) AB14W060100150, UDI/DI 00763000758288, Serial Numbers: B748690, B759617, B762803, B769628, B781383, B792612; 18) AB14W060120150, UDI/DI 00763000758301, Serial Numbers: B762187, B766265, B789926; 19) AB14W060150150, UDI/DI 00763000758325, Serial Numbers: B730835, B734241, B757655, B758869, B761533, B776389, B781300, B791875; 20) AB14W060200150, UDI/DI 00763000758349, Serial Numbers: B726348, B736196, B749504, B750650, B754900, B761765, B762193, B769612, B769616, B769622, B769634, B781989, B792379.
510(k) numbers
["K141118"]
PMA numbers
not on file
Product code
LIT
Firm FEI
2183870
Firm city
Plymouth
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register