FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-00502F, K05-00587D, K05-01052, K05-01706C, K05-01880K, K05-02130, K05-02595, K09-12599A, K10-05487P, K12-00264, MAP111/F, MAP112/B, MAP112/F, K08-YP085A, K08-YS017B, K08-YS055A, K09-YS029A, K09-YS040F, K12-YP002A, K12-YP030A, K12-YP060A, K12-YP061A, K12-YP063C, K12-YP072, K12-YS018A, K12-YS019A, K12-YS026A, K05-01140D, K05-01706C***, K05-02595, MAP112, K05-00564L, K12-YP002A

Recall number
Z-0903-2022
Recalling firm
Merit Medical Systems, Inc.
Product
Access-9 Large Bore Hemostasis Valves, Catalog Code: 580775001, IN8112/C, K05-00053H, K05-00109G, K05-00137K, K05-00410G, K05-00438G, K05-00502F, K05-00587D, K05-01052, K05-01706C, K05-01880K, K05-02130, K05-02595, K09-12599A, K10-05487P, K12-00264, MAP111/F, MAP112/B, MAP112/F, K08-YP085A, K08-YS017B, K08-YS055A, K09-YS029A, K09-YS040F, K12-YP002A, K12-YP030A, K12-YP060A, K12-YP061A, K12-YP063C, K12-YP072, K12-YS018A, K12-YS019A, K12-YS026A, K05-01140D, K05-01706C***, K05-02595, MAP112, K05-00564L, K12-YP002A,
Status
Open, Classified
Initiated
2022-03-01
Posted
not on file
Terminated
not on file
Reason
A design change made to a hemostasis valve to improve manufacturability caused an internal gap between the rotator and the Y-body, where the guidewire can get caught, resulting in difficulty advancing the guidewire through the device. This defect may result in a delay of procedure.
Root cause
Component change control
Action
On 03/01/22, recall notices were mailed to customers wo were informed to take the following actions: 1. Quarantine and discontinue use, and distribution of affected devices. 2. Ensure that applicable personnel within their organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, ensure the recall notice is immediately shared with them and note the quantity distributed on the Customer Response Form. 4. Fill out, scan, and email the completed Customer Response Form to Customer Service at response@merit.com. All affected product shipped to you must be accounted for on the Customer Response Form. 5. Immediately return all affected lots to the recalling firm. Customers with additional questions are encouraged to contact a sales representative or the firm s customer service at RESPONSE-EMEA@merit.com, 1-800-356-3748, Hours: 6 am to 6 pm MST, Mon-Fri. Recall notices were also sent to Sales representatives who were asked to contact their affected accounts and assist with product return and completion of the Customer Response Form. On 03/21/22, recall notices and response forms including an expanded scope were sent to customers informing them of additional distributed amounts of affected devices.
Quantity
36445
Distribution
US: MN, PA, AZ, FL, IL, AK, NY, CA, TX, KY, WI, NC, NJ, VA, CO, IN, AR, MO, LA, OR, MT, MA, MI, NE, WV, KS, MS, HI, IA, NH, ME, AL, DC, DE, NM, OH, GA, MD, TN, UT, RI, SC, OK, WA, ID, ND, SD, NV, VT. OUS: Poland, Germany, France, United Kingdom, Belgium, Korea, Republic of, Austria, Canada, Spain, Luxembourg, Italy, Switzerland, Czech Republic, Netherlands, Japan, Honduras, Colombia, Greece, Ireland
Lot, serial or model codes
Catalog Code/Lots: 580775001/I2219771, I2278741, I2226790; IN8112/C/H2292357, H2287440, H2281404, H2298626, H2305047, H2305048, H2309356; K05-00053H/H2305217; K05-00109G/H2304656; K05-00137K/H2304653; K05-00410G/H2305216; K05-00438G/H2309130; K05-00502F/H2309122; K05-00587D/H2309211; K05-01052/H2304607; K05-01706C/H2309133; K05-01880K/H2306962; K05-02130/H2304639; K05-02595/H2304614; K09-12599A/H2295450; K10-05487P/H2279857; K12-00264/H2285040; MAP111/F/I2259670, I2279884, I2259657, I2289610; MAP112/B/P1908667; MAP112/F/I2259661, I2279915, I2279894, I2176725, I2215955, I2289611, I2279895, I2289613, I2289612; K08-YP085A/H2249051, H2302394; K08-YS017B/H2249052; K08-YS055A/H2275990; K09-YS029A/H2266470, H2302474; K09-YS040F/H2286601; K12-YP002A/H2269321, H2278279, H2283742, H2295661; K12-YP030A/H2263653; K12-YP060A/H2272013; K12-YP061A/H2249088; K12-YP063C/H2272382, H2295556; K12-YP072/H2266224; K12-YS018A/H2284956; K12-YS019A/H2302506; K12-YS026A/H2276014, H2302507; K05-01140D/H2304611; K05-01706C/ (H2304610, H2309133)***; K05-02595/H2304614; MAP112/I2279895; K05-00564L/H2304617; K12-YP002A/H2295661;
510(k) numbers
not on file
PMA numbers
not on file
Product code
DTL
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register