Skip to the content

FDA Medical Device Recalls (openFDA)

Merit Medical Systems, Inc.: Radial Assist RAD BOARD, used to support the weight of a patient's arm and supplies for a medical procedure. The Merit RAD BOARD is a rigid and stiff body board intended for use for various medical purposes. RAD BOARD was specifically designed to support the weight of a patient's arm and supplies for medical procedure, in order to have optimal access to upper extremity vasculature, including radial and brachial arterial and venous access. The RAD BOARD is partially lined with a layer of lead-free Xenolite TB for additional radiation scatter protection.

Recall number
Z-0901-2014
Recalling firm
Merit Medical Systems, Inc.
Product
Radial Assist RAD BOARD, used to support the weight of a patient's arm and supplies for a medical procedure. The Merit RAD BOARD is a rigid and stiff body board intended for use for various medical purposes. RAD BOARD was specifically designed to support the weight of a patient's arm and supplies for medical procedure, in order to have optimal access to upper extremity vasculature, including radial and brachial arterial and venous access. The RAD BOARD is partially lined with a layer of lead-free Xenolite TB for additional radiation scatter protection.
Status
Terminated
Initiated
2013-12-18
Posted
2014-01-30
Terminated
2014-02-13
Reason
Merit Medical Systems, Inc. is voluntarily recalling one lot (B507171) of RAD BOARD RB 100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There
Root cause
Labeling Change Control
Action
Merit Medical Systems, Inc.sent a letter voluntarily recalling one lot (B507171) of RAD BOARD" RB100 devices. The affected devices are missing the main label which graphically depicts patient orientation on the board. The issue may allow the board to be oriented in the wrong direction where the semi-radiopaque material in the board (embedded Xenolite) could obscure patient anatomy during fluoroscopy. There have been no reports of patient harm or injury from Merit customers as a result of this issue. Merit has chosen to remove this lot from the field by replacing the unlabeled RAD BOARDs. Merit requests that customers stop using these devices immediately until the product has been replaced by their Merit sales representative. Customers with questions were instructed to call 1-801-208-4344 or 1-801-208-4365. For questions regarding this recall call 801-208-4301.
Quantity
11
Distribution
Worldwide Distribution - USA including NJ, TX, WV, MN, GA, and Internationally to Mexico and Japan.
Lot, serial or model codes
lot B507171
510(k) numbers
not on file
PMA numbers
not on file
Product code
BWN
Firm FEI
1721504
Firm city
South Jordan
Firm state
UT
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register