Zimmer, Inc.: Osteobond Bone Cement Bone cement monomer is kitted with bone cement powder to create a cement mantle for orthopedic implants.

Recall number
Z-0900-2015
Recalling firm
Zimmer, Inc.
Product
Osteobond Bone Cement Bone cement monomer is kitted with bone cement powder to create a cement mantle for orthopedic implants.
Status
Terminated
Initiated
2014-12-08
Posted
2014-12-31
Terminated
2015-07-17
Reason
Samples of product from Supplier lot 14D0809 and 14D0808 showed little adhesive transfer between the Tyvek lid and cavity when opened for internal bone cement cure and pairing tests. Visually these samples do not meet the requirements of ZWI 43.109 and corresponding visual aid 55-0000-310-01.
Root cause
Employee error
Action
"Urgent Medical Device Recall - Lot Specific" notifications will be sent to the affected distributors, surgeons, and hospital risk managers with instructions for returning the affected product. The letters identified the affected product, reason for recall, risks, and responsibilities. Customers are to complete and return the Inventory Return Certification Form along with the affected product. Questions should be directed to the customer call at 1-800-348-2759.
Quantity
40 units
Distribution
Distributed to MI, KS, and CA.
Lot, serial or model codes
PMA #N17755, decision date 5/17/1976, reclassified 10/14/1999. Class II device; D033511, Item No. 00-1101-012-00; lot 62849791, 14D0809, Manufactured between 9/11/2014-10/6/2014; EXP date: 4/30/2017
510(k) numbers
not on file
PMA numbers
not on file
Product code
LOD
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register