Skip to the content

FDA Medical Device Recalls (openFDA)

McKesson Provider Technologies - Medical Imaging Group: Horizon Cardiology ECG Management This is a software application designed for use in various hospital departments to import, display, store, analyze, distribute, and process ECG procedures from resting ECG devices. It is also intended to provide analysis or reanalysis or testings ECGs and to provide preliminary data for editing and confirmation by an over-reading physician. Horizon Cardiology ECG Management may provide a serial comparison of ECG Data to facilitate the review of the patient's current ECG with previous ECG's of the same patient.

Recall number
Z-0900-2011
Recalling firm
McKesson Provider Technologies - Medical Imaging Group
Product
Horizon Cardiology ECG Management This is a software application designed for use in various hospital departments to import, display, store, analyze, distribute, and process ECG procedures from resting ECG devices. It is also intended to provide analysis or reanalysis or testings ECGs and to provide preliminary data for editing and confirmation by an over-reading physician. Horizon Cardiology ECG Management may provide a serial comparison of ECG Data to facilitate the review of the patient's current ECG with previous ECG's of the same patient.
Status
Terminated
Initiated
2010-06-16
Posted
2011-01-13
Terminated
2012-08-17
Reason
When importing ECG files from Mortara ECG carts using DICOM interface, the Horizon Cardiology ECG Management System does not display the amplitudes of the waveform properly.
Root cause
Device Design
Action
McKesson Medical sent an advisory notice letter dated June 16, 2010, to all affected customers. Advisory notice ADV01646 was distributed to 13 affected clinical sites from August 30, 2010 to September 28, 2010. The letters identified the product that is affected, the circumstances under which patient safety issue may occur, the risk, and the immediate actions that need to be taken along with a temporary fix. Customers were instructed to use the ECG cart printout to verify results of procedures acquired by Mortara ECG carts. Users should be aware of the problem when referrring to past ECG procedures and when performing a visual comparison with ECG procedures acquired by Mortara ECG carts. Customers should disable the Glasgow Serial Comparison for ECG procedures. McKesson Support would contact the customer to coordinate the deployment of the software update on all the affected applications and Horizon Cardiology client applications. For any questions regarding this recall, customers should call the Customer Support department at 1-866-777-0202 and speak to their Support Manager.
Quantity
16 units
Distribution
Nationwide Distribution including AZ, CA, FL, GA, MA, PA, UT, and WI.
Lot, serial or model codes
K061905; Versions 12, 12 SP1, 12 SP1 HF2, 12 SP1 HF4, 12.1, 12.1 HF1, 12.1 HF2
510(k) numbers
not on file
PMA numbers
not on file
Product code
DQK
Firm FEI
3005157178
Firm city
Parsippany
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register