Skip to the content

FDA Medical Device Recalls (openFDA)

Teleflex Medical: Mucosal Atomization Device (MAD), Intranasal Mucosal Atomization Device with 3mL Syringe and Vial Adapter; Product Code MAD100, MAD100OS MAD110, MAD110OS, MAD130, MAD130OS, MAD140, MAD140OS, MAD300, MAD300B. Manufactured By: Wolfe-Tory Medical, Inc., Teleflex Medical. For atomization of topical solutions across the naso and oropharyngeal mucosal membranes

Recall number
Z-0898-2017
Recalling firm
Teleflex Medical
Product
Mucosal Atomization Device (MAD), Intranasal Mucosal Atomization Device with 3mL Syringe and Vial Adapter; Product Code MAD100, MAD100OS MAD110, MAD110OS, MAD130, MAD130OS, MAD140, MAD140OS, MAD300, MAD300B. Manufactured By: Wolfe-Tory Medical, Inc., Teleflex Medical. For atomization of topical solutions across the naso and oropharyngeal mucosal membranes
Status
Terminated
Initiated
2016-10-27
Posted
2016-11-09
Terminated
2020-09-25
Reason
Product Defect; These products may produce a straight stream instead of a fully atomized plume of medication.
Root cause
Process control
Action
Teleflex sent an Urgent Medical Device Recall Notification letter on/about October 27, 2016. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately discontinue use and quarantine affected products. Customers were asked to complete the enclosed Recall Acknowledgement Form and fax it to Teleflex Customer Service or email to recalls@teleflex.com. For questions, feel free to contact your local sales representative or Customer Service at 1-866-246-6990. Teleflex sent an Amended Urgent Medical Device Recall Notification letter dated November 22, 2016, to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to continue to return affected product for a full refund per the procedure set forth in the letter of October 27, 2016. Some customers have indicated that, due to medical necessity, they plan to continue using the affected lots rather than return them. Teleflex is advising all such customers to follow the supplemental instructions for Use included as Attachment B. These supplemental instructions allow non-destructive testing of each unit prior to the procedure to determine if it is defective. Institutions that intend to continue using affected lots with this supplemental testing, please send the acknowledgment form to recall@teleflex.com with their name and contact information and the lots that they will continue to use, or call 866-246-6990. Customers with questions should contact their local sales representative or Customer Service at 1-866-246-6990.
Quantity
2,674,650 units in total
Distribution
Worldwide distribution. U.S. (Nationwide) and Internationally to Canada, Belgium, Singapore and Thailand.
Lot, serial or model codes
Product code MAD100, Batch/Lot numbers 160105, 160137, 160302, 160321, 160402, 160435, 160506, 160523, 160609, 160620, 160707, 160802, 160813;  Product code MAD100OS, Batch/Lot numbers 160322, 160524, 160630;  Product code MAD110, Batch/Lot numbers 160217, 160507; Product code MAD110OS, Batch/Lot numbers 160240, 160312; Product code MAD130, Batch/Lot numbers 160107, 160138, 160517; Product code MAD130OS, Batch/Lot numbers 160436, 160803; Product code MAD140, Batch/Lot numbers 160125, 160218, 160437, 160610, 160801; Product code MAD140OS, Batch/Lot numbers 160226, 160438, 160727; Product code MAD300, Batch/Lot numbers 160108, 160117, 160126, 160145, 160146, 160200, 160219, 160225, 160231, 160300, 160313, 160327, 160400, 160409, 160422, 160432, 160440, 160500, 160518, 160602, 160611, 160621, 160631, 160701, 160708, 160718, 160728, 160800, 160804, 160814, 160816, 160823; Product code MAD300B, Batch/Lot number 160410.
510(k) numbers
not on file
PMA numbers
not on file
Product code
CCQ
Firm FEI
3005747797
Firm city
Research Triangle Park
Firm state
NC
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register