Skip to the content

FDA Medical Device Recalls (openFDA)

Smiths Medical ASD, Inc.: Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 10mL Luer lock syringe; Reorder Number: 4102010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intended for injection or aspiration of fluids utilizing a luer lock or luer slip syringe.

Recall number
Z-0898-2015
Recalling firm
Smiths Medical ASD, Inc.
Product
Hypodermic Needle Pro Edge" Safety Device with Syringe 20G x1 10mL Luer lock syringe; Reorder Number: 4102010 Product Usage: The Needle Pro Edge Safety Hypodermic Needle is intended for injection or aspiration of fluids utilizing a luer lock or luer slip syringe.
Status
Completed
Initiated
2014-11-21
Posted
2014-12-30
Terminated
not on file
Reason
Needle is not captured in the needle safety sheath in specific lots.
Root cause
Process control
Action
Smith Medical sent on 11/24/2014, consignees an Urgent Field Safety Notice via FedEx Overnight. The notification instructs customers to inspect their inventory and complete and return the attached Urgent Field Safety Notice Confirmation Form to coordinate return of affected product to Smiths Medical for replacement/credit. For questions regarding this information, please contact Smiths Medicals Customer Service Department at 800-258-5361.
Quantity
47,200
Distribution
US Nationwide Distribution in the states of AZ CA FL GA IL IN LA MA MN MO MS MT NC ND NJ NV NY OH OR PA RI TN TX VA WA WI WV
Lot, serial or model codes
Lot Numbers: 2756526, 2771498
510(k) numbers
["K041399"]
PMA numbers
not on file
Product code
FMI
Firm FEI
1217052
Firm city
Keene
Firm state
NH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register