Skip to the content

FDA Medical Device Recalls (openFDA)

Arizant Inc: Electri-Cord Manufacturing Corporation (312 E. Main St., Westfield, PA 16950) AC power cords used with Ranger Irrigation Fluid Warming Unit; Model 247. Intended to warm irrigation fluids.

Recall number
Z-0897-2010
Recalling firm
Arizant Inc
Product
Electri-Cord Manufacturing Corporation (312 E. Main St., Westfield, PA 16950) AC power cords used with Ranger Irrigation Fluid Warming Unit; Model 247. Intended to warm irrigation fluids.
Status
Terminated
Initiated
2009-12-17
Posted
2010-03-01
Terminated
2012-04-19
Reason
Arizant Healthcare Inc. is initiating a nationwide voluntary recall of 100/110-volt power cords (Arizant part number 502221 Model 90024) attached to its temperature management units and pressure infusion power pack (Bair Paws Temperature Management Units: Models 850 and 875 Bair Hugger Temperature Management Units: Models 500/OR, 505 (Human & Vet.), 750 (Human & Vet) and 775 Ranger Blood/Flui
Root cause
Nonconforming Material/Component
Action
An "Urgent Medical Device Power Cord Recall" letter dated December 17, 2009 was sent to customers. The letter was addressed to Recall Coordinator/Biomed Dept. The letter described the affected product, problem, hazard involved, and action to be taken by customer. The customers are being ask to inspect their temperature management units (as described in FDA's recommendation dated October 19, 2009) and contact the firm with the results of their inspection and provide a contact name via website www.arizant.com/powercords or phone (877) 947-1487. Replacement cords and instructions will be provided to customers based on priority listed in the letter and availability of replacement cords. If the customer does not have the means to make the changes, alternate means will be taken (i.e., an Arizant employee or outside service organizations may be brought in to do the replacements). If you have any questions, please contact the Aziant at (877) 47-1487.
Quantity
Total distributed for all devices: 84,537 US, 7,380 OUS.
Distribution
Worldwide Distribution: USA and ARGENTINA, THE BAHAMAS, CANADA, COLOMBIA, COSTA RICA, CUBA, DOMINICAN REPUBLIC, ECUADOR, GERMANY, GUAM, JAPAN, LEBANON, MEXICO, NICARAGUA, PANAMA, PUERTO RICO, QATAR, SAUDI ARABIA, SOUTH KOREA, SURINAME , TAIWAN, TRINIDAD AND TOBAGO, QATAR.
Lot, serial or model codes
Part number 502221 Model 90024
510(k) numbers
["K053645","K060865"]
PMA numbers
not on file
Product code
DWJ
Firm FEI
2183725
Firm city
Eden Prairie
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register