FDA Medical Device Recalls (openFDA)
Arizant Inc: Electri-Cord Manufacturing Corporation (312 E. Main St., Westfield, PA 16950) AC power cords used with Ranger Irrigation Fluid Warming Unit; Model 247. Intended to warm irrigation fluids.
- Recall number
- Z-0897-2010
- Recalling firm
- Arizant Inc
- Product
- Electri-Cord Manufacturing Corporation (312 E. Main St., Westfield, PA 16950) AC power cords used with Ranger Irrigation Fluid Warming Unit; Model 247. Intended to warm irrigation fluids.
- Status
- Terminated
- Initiated
- 2009-12-17
- Posted
- 2010-03-01
- Terminated
- 2012-04-19
- Reason
- Arizant Healthcare Inc. is initiating a nationwide voluntary recall of 100/110-volt power cords (Arizant part number 502221 Model 90024) attached to its temperature management units and pressure infusion power pack (Bair Paws Temperature Management Units: Models 850 and 875 Bair Hugger Temperature Management Units: Models 500/OR, 505 (Human & Vet.), 750 (Human & Vet) and 775 Ranger Blood/Flui
- Root cause
- Nonconforming Material/Component
- Action
- An "Urgent Medical Device Power Cord Recall" letter dated December 17, 2009 was sent to customers. The letter was addressed to Recall Coordinator/Biomed Dept. The letter described the affected product, problem, hazard involved, and action to be taken by customer. The customers are being ask to inspect their temperature management units (as described in FDA's recommendation dated October 19, 2009) and contact the firm with the results of their inspection and provide a contact name via website www.arizant.com/powercords or phone (877) 947-1487. Replacement cords and instructions will be provided to customers based on priority listed in the letter and availability of replacement cords. If the customer does not have the means to make the changes, alternate means will be taken (i.e., an Arizant employee or outside service organizations may be brought in to do the replacements). If you have any questions, please contact the Aziant at (877) 47-1487.
- Quantity
- Total distributed for all devices: 84,537 US, 7,380 OUS.
- Distribution
- Worldwide Distribution: USA and ARGENTINA, THE BAHAMAS, CANADA, COLOMBIA, COSTA RICA, CUBA, DOMINICAN REPUBLIC, ECUADOR, GERMANY, GUAM, JAPAN, LEBANON, MEXICO, NICARAGUA, PANAMA, PUERTO RICO, QATAR, SAUDI ARABIA, SOUTH KOREA, SURINAME , TAIWAN, TRINIDAD AND TOBAGO, QATAR.
- Lot, serial or model codes
- Part number 502221 Model 90024
- 510(k) numbers
- ["K053645","K060865"]
- PMA numbers
- not on file
- Product code
- DWJ
- Firm FEI
- 2183725
- Firm city
- Eden Prairie
- Firm state
- MN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28