Skip to the content

FDA Medical Device Recalls (openFDA)

Hansen Medical Inc: Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.

Recall number
Z-0896-2015
Recalling firm
Hansen Medical Inc
Product
Sensei X Robotic Catheter System, model numbers 02057 and 08257. - Set up joint (SUJ)mechanical clamp that secures the SUJ to the catheter laboratory table. Product Usage: The Sensei X Robotic Catheter System is an electromechanical software controlled system designed to facilitate navigation and control of Hansen Medicals robotically steerable catheters for collecting electrophysiological data within the heart atria with electro-anatomic mapping and recording systems.
Status
Terminated
Initiated
2014-12-04
Posted
2014-12-30
Terminated
2015-09-03
Reason
The Set Up Joint subassembly of the Sensei X Robotic Catheter System may become loose relative to the operating table with the potential for the device to fall causing, uncontrolled catheter movement. cardiac tampenade, delayed procedure, or crush injury to the patient and/or user.
Root cause
Device Design
Action
Hansen Medical sent an Urgent Notice - Field Safety Notice letters to US customers on December 4, 2014 via Federal Express. The letter identified the affected product, problem and actions to be taken. Customers were instructed to complete and fax back the attached form to Hansen medical at 650-404-2773. For further information call 650-404-5841.
Quantity
59 USA, 40 OUS
Distribution
Worldwide Distribution - US Nationwide and the countries of Italy, Czech Republic, Russia, Spain, Singapore, France, Germany, Denmark, Netherlands and Australia.
Lot, serial or model codes
Model numbers 02057 and 08257, all installed Sensei X Robotic systems are subject to this action.
510(k) numbers
["K052480","K073225","K091808","K102168","K141822"]
PMA numbers
not on file
Product code
DXX
Firm FEI
3006026430
Firm city
Mountain View
Firm state
CA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28