FDA Medical Device Recalls (openFDA)
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING: GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission
- Recall number
- Z-0894-2025
- Recalling firm
- GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
- Product
- GE Healthcare Discovery NM/CT 670 ES, Model/Catalog Numbers: 1) H3100XF ; System, Tomography, Computed, Emission
- Status
- Open, Classified
- Initiated
- 2024-12-18
- Posted
- 2025-01-10
- Terminated
- not on file
- Reason
- GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.
- Root cause
- Package design/selection
- Action
- GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 12/18/2024. The notice explained the safety issue and requested the following actions be taken: There are two populations potentially affected by this notification: NOTE: Specific serial numbers of the following affected systems: Infinia Nuclear Medicine Systems: Infinia, Infinia Hawkeye, Infinia II, Infinia II Hawkeye 4; VG Nuclear Medicine systems: Varicam, Millennium VG, Millennium VG Hawkeye, Discovery VH; Brivo NM 615; Discovery NM 630; Discovery NM/CT 670, Discovery NM/CT 670 ES, Discovery NM/CT 670 DR, Discovery NM/CT 670 PRO, Discovery NM/CT 670 CZT; Optima NM/CT 640. For Serial Numbers listed in Appendix A of this letter: GE HealthCare requests that you STOP USING the system until GE HealthCare has inspected and if necessary, corrected your system and provided you with a letter indicating the system can be returned to clinical use. GE HealthCare has determined that these systems could have previously been transported or relocated without proper detector support. For Serial Numbers listed in Appendix B of this letter: You can continue to use your system, as GE HealthCare has determined that these systems have not previously been transported or relocated without proper detector support. For Serial Numbers listed in either Appendix A or Appendix B of this letter: GE HealthCare requests that you DO NOT physically relocate or transport your system from its current location. If you need to move or transport your system, please email GE HealthCare at recall.40910@gehealthcare.com to receive specific instructions to follow before moving the system. When emailing this address, please provide the following information: " System serial number " Current system location " The reason for the move or transport, and subsequent destination Do not proceed with a physical relocation or transport of your system until you have received these instructions and taken the actions descri
- Quantity
- 17 units
- Distribution
- Worldwide
- Lot, serial or model codes
- 1) H3100XF: UDI/DI N/A, System ID NT1011, Serial Number T2AG12009/ 88227HL4; UDI/DI N/A, System ID NT2002, Serial Number T2AG12011/ 89154HL9; UDI/DI N/A, System ID NT1009, Serial Number T2AG12013/ 90088HL6; UDI/DI N/A, System ID NT1003, Serial Number T2BG11001/ 79573HL2; UDI/DI N/A, System ID NT1004, Serial Number T2AG12003/ 79574HL0; UDI/DI N/A, System ID NT1005, Serial Number 84775HL6; UDI/DI N/A, System ID NT1020, Serial Number T2AG12012/ 89666HL2; UDI/DI N/A, System ID NT1016, Serial Number T2AG13020/ 96237HL3; UDI/DI N/A, System ID NT0002, Serial Number 106491HL4; UDI/DI N/A, System ID NT2004, Serial Number 105724HL9; UDI/DI N/A, System ID NT1012, Serial Number T2AG13016/ 93885HL2; UDI/DI N/A, System ID NT1012, Serial Number T2AG13017/ 93885HL2; UDI/DI N/A, System ID NT1013, Serial Number 94165HL8; UDI/DI N/A, System ID NT1010, Serial Number T2AG12008/ 87697HL9; UDI/DI N/A, System ID NT3001, Serial Number 85748HL2; UDI/DI N/A, System ID NT1017, Serial Number T2AG13027/ 104213HL4; UDI/DI N/A, System ID NT1015, Serial Number T2AG13018/ 94292HL0.
- 510(k) numbers
- ["K093514"]
- PMA numbers
- not on file
- Product code
- KPS
- Firm FEI
- 3004571294
- Firm city
- Tirat Carmel
- Firm state
- not on file
- Firm country
- Israel
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28