Ventana Medical Systems Inc: BenchMark ULTRA and DISCOVERY ULTRA Instruments

Recall number
Z-0894-2022
Recalling firm
Ventana Medical Systems Inc
Product
BenchMark ULTRA and DISCOVERY ULTRA Instruments
Status
Open, Classified
Initiated
2022-01-20
Posted
2022-04-08
Terminated
not on file
Reason
Potential for Fluid Leak inside a staining system that could cause an electrical short circuit of the General Purpose Input/ Output Board's J1 Connector that could result in smoke and fire.
Root cause
Device Design
Action
On 01/20/22, correction notices were sent to customers who were asked to take the following actions needed to continue affected device use: 1) Inform any operators of the potential hazards associated with this issue and provide a copy of this notification as appropriate. 2) Inspect the instrument daily for evidence of leaks (i.e., liquid on the floor, or in and around the instrument). 3) If you experience a fluid leak, stop using the system and turn off the power to the instrument. Immediately contact the firm's Network Customer Support Center at 1-800-227-2155. 4) Complete the attached fax back form and return via fax or email. 5) If your facility has distributed the affected product to another site, please ensure this UMDC is provided to that site. 6) File this UMDC for future reference. The firm will send a representative to schedule a visit to perform an inspection, reroute cables, and replace parts as necessary to help mitigate the issue. Customers with questions were encouraged to call the Customer Support Center at the before mentioned number; support is available 24 hours a day, seven days a week. On 4/19/23, correction notices were distributed to additional customers who may have received devices without a drip loop, which is the solution to this event. Customers were told they would be contacted by a Service Representative to coordinate specific service actions, including the inspection of GPIO boards and rerouting cables - drip loop introduction.
Quantity
9214
Distribution
US: CA, IL, PA, FL, AZ, NY, OH, CO, AR, NC, MD, AL, AK, LA, IA, MN, MA, ND, IN, TX, TN, CT, SC, GA, UT, WI, VA, SD, KY, MS, MO, DE, MI, MT, NJ, WV, OK, WA, OR, PR, DC, NM, ID, NH, NE, ME, KS, HI, NV, RI, WY, VT OUS: Australia, Austria, Belgium, Canada, China, Croatia, Denmark, Finland, France, Germany, Hong Kong, Hungary, India, Israel, Italy, Japan, Netherlands, Norway, Poland, Portugal, Russian Fed., Singapore, South Korea, Spain, Sweden, Switzerland, United Kingdom, Algeria, Argentina, Aruba, Bosnia-Herzegovina., Brazil, Bulgaria, Chile, Colombia, Costa Rica, Cuba, Cyprus, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Greece, Indonesia, Ireland, Kazakhstan, Kuwait, Latvia, Lebanon, Lithuania, Luxembourg, Macedonia, Malaysia, Martinique, Mexico, Monaco, Morocco, Namibia, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Reunion, Romania, Saudi Arabia, Serbia, Slovakia, Slovenia, South Africa, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, Vietnam, White Russia, Zimbabwe
Lot, serial or model codes
Serial Numbers Less-Than-Or-Equal-To 322030
510(k) numbers
not on file
PMA numbers
not on file
Product code
KPA
Firm FEI
1000125183
Firm city
Oro Valley
Firm state
AZ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register