FDA Medical Device Recalls (openFDA)
Abbott Health Products, Inc.: IMx total B-hCG Reagent is composed of one bottle of Coated Microparticles, one bottle of Alkaline Phosphatase Conjugate, one bottle of 4 Methylumbelliferyl Phosphate and one bottle of Specimen Diluent
- Recall number
- Z-0894-05
- Recalling firm
- Abbott Health Products, Inc.
- Product
- IMx total B-hCG Reagent is composed of one bottle of Coated Microparticles, one bottle of Alkaline Phosphatase Conjugate, one bottle of 4 Methylumbelliferyl Phosphate and one bottle of Specimen Diluent
- Status
- Terminated
- Initiated
- 2005-03-24
- Posted
- 2005-06-16
- Terminated
- 2005-09-29
- Reason
- Abbott Diagnostics is investigating a recent increase in customer complaints for IMx calibration error code 157 and controls out of range when using IMx total B-hCG Reagent Lot 25459Q100. Testing has shown that protein aggregates are present in the specimen diluents of both IMx and the AxSYM Total B-hCG reagent kits. This is responsible for the calibration errors and shift in control values.
- Root cause
- Other
- Action
- Abbot Health Products, Inc informed and instructed their customers via Device Recall Letter to Identify if they have used or have inventory of the affected product. Discontinue use of and destroy any remaining inventory of reagent lots listed in the Device Recall letter according to their laboratory procedures. Record the quantity of the recalled reagents currently in inventory in their facility that will be destroyed on the attached reply form and fax the form back to Abbott 1-800-777-0051. (US only) Follow their laboratory''s procedure for communicating this issue to the health care providers they serve.
- Quantity
- 793 kits
- Distribution
- US Goverment, Domestic and International distribution.
- Lot, serial or model codes
- 25459Q100
- 510(k) numbers
- ["K912535"]
- PMA numbers
- not on file
- Product code
- JHI
- Firm FEI
- 2623532
- Firm city
- Barceloneta
- Firm state
- not on file
- Firm country
- PR
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28