International Technidyne Corp.: Hemochron Signature Elite Whole Blood Microcoagulation (IVD) System. The system is intended to be used only with test cuvettes that are available from ITC.

Recall number
Z-0893-2011
Recalling firm
International Technidyne Corp.
Product
Hemochron Signature Elite Whole Blood Microcoagulation (IVD) System. The system is intended to be used only with test cuvettes that are available from ITC.
Status
Terminated
Initiated
2010-03-31
Posted
2011-03-01
Terminated
2012-09-25
Reason
There is a mismatch between the class of the laser barcode scanner and the labeling of the Hemochron Signature Elite: some of the devices were inadvertently equipped with class II laser barcode scanner instead of the intended class I barcode scanner.
Root cause
Labeling mix-ups
Action
Urgent - Medical Device Recall - Hemochron Signature Elite letters, dated February 24, 2010, by Federal Express to all direct customers. The letter informed customers of the issue and stated that the problem does not negatively affect the performance of the product and to continue using the product as intended. In addition, customers are to notify the appropriate personnel about the notice, complete and return the attached form, and an ITC representative will contact customers to discuss arrangements to re-label their devices. Customers can contact ITC Technical Support at 800-579-2255 (US) or 732-548-5700 ext 4700 (International) if they have any questions regarding this issue and notice. Customers can also email sigelitescan@itcmed.com with questions.
Quantity
1640 units (1303 US, 337 foreign)
Distribution
Worldwide Distribution -- United Arab Emirates, Belgium, Canada, Switzerland, Germany, Denmark, France, United Kingdom, Ireland, Iceland, Italy, Kuwait, Liechtenstein, Netherlands, Oman, Qatar, Saudi Arabia, Sweden, and Singapore.
Lot, serial or model codes
Serial numbers SE2830 to SE2854 (inclusive); SE2880 to SE4789 (inclusive); SE0260, SE0366, SE0567, SE0764, SE0843, SE0970, SE1086, SE1198, SE1208, SE1212, SE1438, SE1456, SE1580, SE1711, SE2029, SE2136, SE2602, and SE2618.
510(k) numbers
not on file
PMA numbers
not on file
Product code
RFZ
Firm FEI
1000526865
Firm city
Edison
Firm state
NJ
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register