Synovis Surgical Inovation Div. of: Peri-Strips Dry with Veritas Collagen Matrix Staple Line Reinforcement (PSD-V), 4506 E-V. PSD-V is an implantable surgical mesh comprised of non-crosslinked bovine pericardium in a dehydrated form. Do not reuse. Consult Instructions for Use. Sterilized. Product treated with sodium hydroxide. Produt derived from USDA-inspected cattle. Made in the U.S.A. Rx Only. Synovis Surgical Innovations A Division of Synovis Life Technologies, Inc., 6575 University Ave W., St. Paul, MN 55114-1024 USA. 651 796-7300, 800.487.9627

Recall number
Z-0892-05
Recalling firm
Synovis Surgical Inovation Div. of
Product
Peri-Strips Dry with Veritas Collagen Matrix Staple Line Reinforcement (PSD-V), 4506 E-V. PSD-V is an implantable surgical mesh comprised of non-crosslinked bovine pericardium in a dehydrated form. Do not reuse. Consult Instructions for Use. Sterilized. Product treated with sodium hydroxide. Produt derived from USDA-inspected cattle. Made in the U.S.A. Rx Only. Synovis Surgical Innovations A Division of Synovis Life Technologies, Inc., 6575 University Ave W., St. Paul, MN 55114-1024 USA. 651 796-7300, 800.487.9627
Status
Terminated
Initiated
2005-05-23
Posted
2005-06-14
Terminated
2005-11-17
Reason
Device is contained in 2 pouches. The outer pouch's seal may be compromised; this may render the exterior of the interior pouch non-sterile.
Root cause
Other
Action
The sites and/or distributors that received the suspect devices were contacted by telephone and given instruction how to return/replace the devices. A Recall Notice, dated 05/31/05, was also sent to customers indicating the model number(s) and quantities of affected product which the institution received and the amount they were to return.
Quantity
56 boxes each box containing 6 pouches
Distribution
AK, AL, AZ, CA, DE, FL, GA, MO, OR, TN, TX, VA and WA
Lot, serial or model codes
lot number: 5707471-116932, 5707579-118319 and 5709356-167628.
510(k) numbers
not on file
PMA numbers
not on file
Product code
FTM
Firm FEI
2183620
Firm city
Saint Paul
Firm state
MN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register