Abbott Laboratories: ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.
- Recall number
- Z-0891-2022
- Recalling firm
- Abbott Laboratories
- Product
- ARCHITECT STAT Myoglobin Calibrators, RES 2K43-01 The ARCHITECT STAT Myoglobin Calibrators are for the calibration of the ARCHITECT iSystem with STAT protocol capability when used for the quantitative determination of myoglobin in human serum and plasma.
- Status
- Open, Classified
- Initiated
- 2022-03-08
- Posted
- not on file
- Terminated
- not on file
- Reason
- This calibrator lot did not meet acceptance criteria during ongoing stability testing. When a calibration is performed with this lot, controls may exceed the specified range in the Instructions for Use (IFU).
- Root cause
- Under Investigation by firm
- Action
- The firm, Abbott, notified their consignees by "Urgent Medical Device Products Recall" letter on 03/08/2022. The letter explained the issue and requested the use of the affected item be discontinued and destroyed, and the Customer Reply Form be completed and returned via FAX #: 1-800-777-0051 or email PMS@abbott.com. The firm is replacing the product. If you or any of the health care providers you serve have questions regarding this information, U.S. Customers please contact Customer Service at 1-877-4ABBOTT (available 24 hours a day, 7 days a week).
- Quantity
- 402 kits
- Distribution
- Worldwide Distribution: US (nationwide) and OUS (foreign) including countries of: AUSTRIA, BELGIUM, BOSNIA AND HERZEGOVI, CROATIA, CYPRUS, CZECH REPUBLIC, DENMARK, FRANCE, GERMANY, GREECE, IRAQ, ITALY, KAZAKHSTAN, LITHUANIA, LUXEMBOURG, MACEDONIA, MOROCCO, PANAMA, PORTUGAL, QATAR, ROMANIA, SAUDI ARABIA, SOUTH AFRICA, SWEDEN, SWITZERLAND, and TURKEY
- Lot, serial or model codes
- UDI: (01)00380740003272(17)220531(10)166000
- 510(k) numbers
- ["K042924"]
- PMA numbers
- not on file
- Product code
- JIS
- Firm FEI
- 1415939
- Firm city
- Abbott Park
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28