Skip to the content

FDA Medical Device Recalls (openFDA)

Zimmer Inc.: TiTLE 2 Polyaxial Spinal System, Counter Torque Assembly Tool, Part ET1132-01, Zimmer, Warsaw, IN. The Zimmer Title 2 counter torque assembly tool is a reuseable, non- implantable surgical tool that is provided as part of a the device system for use as a secondary option for compression and distraction of the construct assembly and initial tightening of set screws. The instrument is designed to provide equal torque in the opposite direction to the torque that is being applied to the cap screw during tightening.

Recall number
Z-0887-2011
Recalling firm
Zimmer Inc.
Product
TiTLE 2 Polyaxial Spinal System, Counter Torque Assembly Tool, Part ET1132-01, Zimmer, Warsaw, IN. The Zimmer Title 2 counter torque assembly tool is a reuseable, non- implantable surgical tool that is provided as part of a the device system for use as a secondary option for compression and distraction of the construct assembly and initial tightening of set screws. The instrument is designed to provide equal torque in the opposite direction to the torque that is being applied to the cap screw during tightening.
Status
Terminated
Initiated
2010-10-21
Posted
2011-01-10
Terminated
2011-10-12
Reason
There is a chance that the ball plunger could fall out during surgery and fall into the surgical site. There could be a delay in surgery to remove the ball plunger. If the problem was undetected, the ball plunger could remain in the patient causing pain and potentially necessitating subsequent surgery for removal.
Root cause
Other
Action
Zimmer sent an URGENT MEDICAL DEVICE RECALL LETTER dated October 21, 2010, to all affected customers. The letter identified the product, the problem, the action to be taken by the customer. Customers were instructed to: Inspect their inventory, stop using the device, and place it in quarantine. Contact Zimmer Customer Service at 1-800-774-6368 to obtain a return authorization number. Complete the Inventory Return Certification form and fax a copy to (952) 837-6958, or scan and email to qualitycompliance@zimmer.com. Return the recalled product along with the original completed Return Certification Form to: Zimmer Spine, Inc. Attn: RA Receiving 5301 Riata Park Ct. Austin, Texas 78727 For questions regarding this recall call Zimmer Spine Customer Service at 1-866-774-6368. .
Quantity
68
Distribution
Nationwide Distribution including AL, AZ, CA, CO, FL, MI, MN, NC, PA, OH, OK, TN, TX, WA, and WI.
Lot, serial or model codes
L75118
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1000220733
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register