Skip to the content

FDA Medical Device Recalls (openFDA)

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.: Azurion systems with software release R1.x

Recall number
Z-0886-2022
Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Product
Azurion systems with software release R1.x
Status
Open, Classified
Initiated
2021-05-26
Posted
2022-05-03
Terminated
not on file
Reason
In the Azurion system, the user can add a new study to a patient by selecting the option Add Study . The Add Study dialogue box is then displayed where the Patient Type is selected to perform the study. Due to a software defect, when the study is initiated by pressing Start Procedure , the Patient Type changes inadvertently to a Patient Type different than the one selected as shown in the Table below. Patient type is one of the factors involved in the dose control process. The incorrect patient type changes the technique factors to be used by the system without notification to the user.
Root cause
Software design
Action
Philips notified customers of the issue via Electronic Product Radiation Defect notification letter dated March 4, 2022. The letter identified the affected product, problem and actions to be taken. Customers were informed that they will be contacted by a Philips Field Service Engineers to schedule the software update at no cost. For questions contact your local Philips representative: 1-800-722-9377.
Quantity
722 systems
Distribution
US Nationwide and Worldwide Distribution
Lot, serial or model codes
software release R1.x
510(k) numbers
["K172822"]
PMA numbers
not on file
Product code
OWB
Firm FEI
3002807880
Firm city
Best
Firm state
not on file
Firm country
Netherlands

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register