Fresenius Kabi USA, LLC: Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.

Recall number
Z-0885-2026
Recalling firm
Fresenius Kabi USA, LLC
Product
Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Software for the controlled administration of fluids to patients.
Status
Open, Classified
Initiated
2025-11-14
Posted
2025-12-16
Terminated
not on file
Reason
Software version 5.10.1 and earlier contain anomalies that have the potential to cause serious patient harm or death.
Root cause
Software design
Action
Consignees were emailed an "URGENT - Medical Device Field Correction" notice dated 11/14/25. The notice instructs consignees to review and become familiar with the provided notification's risk mitigation actions; the notice is to be forwarded if affected units were further distributed and users are to be notified. Consignees are to facilitate the installation of the new Ivenix Infusion Management System (version 5.2.2) and Large Volume Pump (5.10.2) software versions utilizing the provided instructions. Consignees are to note that the LVP will not be available for use during the update. Completed Customer Reply Forms may be sent via email to fkmd.fieldactions@fresenius-kabi.com. Customer questions can be directed to Customer Support via email at Ivenix_support@fresenius-kabi.com or by phone at 1-855-354-6387.
Quantity
30 units
Distribution
US (Domestic) distribution: CA, CO, FL, GA, ID, IL, MD, MI, MN, MS, NE, NJ, NV, OK, OR, SC, TX, VA, WA & WI.
Lot, serial or model codes
Software Versions 5.10.1 and prior; UDI: 00811505030122.
510(k) numbers
["K183311"]
PMA numbers
not on file
Product code
FRN
Firm FEI
3014732157
Firm city
North Andover
Firm state
MA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register