Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: HipLOC CHS Plate/Lag Screw Introducer

Recall number
Z-0885-2016
Recalling firm
Biomet, Inc.
Product
HipLOC CHS Plate/Lag Screw Introducer
Status
Terminated
Initiated
2016-02-16
Posted
2016-02-25
Terminated
2017-02-14
Reason
A complaint was received where the Hiploc Plate/Lag Screw Introducer was undersized and would not accept the introducor coupling screw. It appears now that not all parts manufactured to Rev A drawing were reworked.
Root cause
Process change control
Action
Biomet sent an " URGENT MEDICAL DEVICE RECALL NOTICE" dated February 2016, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The following actions are REQUIRED: Immediately locate and remove the identified device(s) listed below from circulation. Remove the affected product from circulation. Review this notification and ensure that all affected personnel are aware of its contents. Carefully follow the instructions on the enclosed Response Form. Email a copy of the response form to CPWARFieldAction@zimmerbiomet.com prior to return of product. You do not need to complete the Biomet UK response form. Use priority carrier for your shipment. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA: "Online: www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or "Call (800)FDA-1088 Thank you in advance for your assistance and prompt attention. On behalf of Zimmer Biomet, I apologize for any inconvenience this may cause. Questions related to this notice should be directed to (574) 372-1570, Monday through Friday, 8 a.m. to 5 p.m.
Quantity
23 units
Distribution
US Distribution to the states of : MO, UT, AZ, IL, PA, KY, MI, NC, NJ, MN and GA., and Internationally to The Netherlands.
Lot, serial or model codes
Catalog Number: 35-000263; and Lot Number Identification: 1807188, 1807189, 1943563, and 2577847.
510(k) numbers
not on file
PMA numbers
not on file
Product code
LXH
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register