FDA Medical Device Recalls (openFDA)
Amendia, Inc: Savannah-T Straight Rod, 200mm, Part Number 10704-200; Savannah-T Straight Rod, 250mm, Part Number 10704-250; Savannah T-Straight Rod 300mm, Part Number 10704-300; Savannah-T Straight Rod, 400mm, Part Number 10704-400; Savannah-T Straight Rod, 500mm; Savannah-T High Top Straight Rod, 200mm; Savannah-T High Top Straight Rod, 220mm, Part Number 1705-220; Savannah-T High Top Straight Rod, 240mm, Product Code 10705-240; Savannah-T Straight Rod, 260mm, Part Number 10705-260; Savannah-T Straight Rod, 280mm, Part Number 10705-280; Savannah-T Straight Rod, 300mm, Part Number 10705-300; Savannah-T Notched Curved Rod, 200mm, Part Number 10706-200; Savannah-T Curve Rod, 200mm, Part Number 10707-200 The guide wire is passed through the cannulated handle of the Dual Stylet Needles and pressed into the bone
- Recall number
- Z-0883-2017
- Recalling firm
- Amendia, Inc
- Product
- Savannah-T Straight Rod, 200mm, Part Number 10704-200; Savannah-T Straight Rod, 250mm, Part Number 10704-250; Savannah T-Straight Rod 300mm, Part Number 10704-300; Savannah-T Straight Rod, 400mm, Part Number 10704-400; Savannah-T Straight Rod, 500mm; Savannah-T High Top Straight Rod, 200mm; Savannah-T High Top Straight Rod, 220mm, Part Number 1705-220; Savannah-T High Top Straight Rod, 240mm, Product Code 10705-240; Savannah-T Straight Rod, 260mm, Part Number 10705-260; Savannah-T Straight Rod, 280mm, Part Number 10705-280; Savannah-T Straight Rod, 300mm, Part Number 10705-300; Savannah-T Notched Curved Rod, 200mm, Part Number 10706-200; Savannah-T Curve Rod, 200mm, Part Number 10707-200 The guide wire is passed through the cannulated handle of the Dual Stylet Needles and pressed into the bone
- Status
- Terminated
- Initiated
- 2016-11-22
- Posted
- not on file
- Terminated
- 2024-02-01
- Reason
- The rods are packaged in packages with unacceptable seals. In some cases, the integrity of the seal is compromised by channels or voids.
- Root cause
- Employee error
- Action
- Amendia sent an Urgent Medical Device Recall letter dated November 22, 2016, to all affected customers. The letter requested that customers examine their inventory and quarantine the product. The letter also requested that they contact Amendia to confirm receipt of the letter and to obtain a Return Material Authorization for any product on hand. Customers with questions were instructed to call the following numbers: 770-575-5224 770-575-5236 770-575-5369 For questions regarding this recall call 404-610-7215.
- Quantity
- 240 units
- Distribution
- Nationwide Distribution to AL, AZ, AR, CA, CO, FL, GA, IA, KS, KY, LA, MA, MI, MN, MO, NV, NJ, NC, SC, TN, TX, VA, WA, WV, WI and Puerto Rico
- Lot, serial or model codes
- Part Number 10704-200 Lot Number 124422, 129766, 130739, 132430; Part Number 10704-250 Lot Number 129790, 130738, 131211, 131309, 131310, 131509, 132811; Part Number 10704-300 Lot Number 129862, 134229; Part Number 10704-400 Lot Number 129861, 130145, 130147, 130502, 131499, 131646; Part Number 10704-500 Lot Number 127934, 127937, 129770, 129863, 130130, 130143, 130781, 134228; Part Number 10705-200 Lot Number 122654, 123583, 129773, 130131, 132153, 132154, 132762; Part Number 10705-220 Lot Number 128818, 131311, 132881; Part Number 10705-240 Lot Number 130500; Part Number 10705-260 Lot Number 121521, 131500, 132199; Part Number 10705-280 Lot Number 121519, 130740; Part Number 10705-300 Lot Number 130140, 130332, 134299; Part Number 10706-200 Lot Number 128522, 129567, 129781, 129891, 130741; Part Number 10707-200 Lot Number 128815, 129782, 129783, 129784, 130182, 131410, 132880.
- 510(k) numbers
- ["K132925","K152920","K153152","K161842"]
- PMA numbers
- not on file
- Product code
- MNH
- Firm FEI
- 3007970248
- Firm city
- Marietta
- Firm state
- GA
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28