Roche Diagnostics Operations, Inc.: ACCU-CHEK Tender I 13/60 10 pieces Product Usage: Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.
- Recall number
- Z-0883-2015
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Product
- ACCU-CHEK Tender I 13/60 10 pieces Product Usage: Usage: Intended for the infusion and/or injection of fluids into the body below the surface of the skin. The indwelling catheter can be inserted independently from the infusion catheter and can be accessed for the injections through the injection port. The indwelling catheter can also be securely attached to the infusion catheter by means of a proprietary click-lock connector for the infusion of drugs subcutaneously. The infusion set can be detached from the indwelling catheter, and the catheter capped to allow freedom from the infusion set and pump for showers, athletics or other activities.
- Status
- Terminated
- Initiated
- 2014-11-20
- Posted
- 2014-12-30
- Terminated
- 2018-01-10
- Reason
- Roche Diabetes Care was informed by UnoMedical, the Manufacturer of ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets, about potential for the the tubing of the infusion set to become detached at the connect/disconnect location on the ACCU-CHEK Tender I and ACCU-CHEK Tender II infusion sets. If tubing detachment occurs, insulin delivery is interrupted and the pump will not alarm to notify
- Root cause
- Device Design
- Action
- Roche sent the "Urgent Medical Device Corrections" letter dated 12/1/2014 via UPS Ground (receipt required) to the Patients, and Distributors who received the affected device. The letter included a description of the reason for recall, affected product, consignee responsibilities, and instructions for responding to the formal recall notification. No product returns were expected. For questions, contact the ACCU-CHEK Pump Support at 1-800-688-4578.
- Quantity
- Total of 13,041 boxes
- Distribution
- US Nationwide Distribution
- Lot, serial or model codes
- Catalog: 04541405001 and Lots: 5044359, 5061929
- 510(k) numbers
- ["K972135"]
- PMA numbers
- not on file
- Product code
- FPA
- Firm FEI
- 1823260
- Firm city
- Indianapolis
- Firm state
- IN
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28