Skip to the content

FDA Medical Device Recalls (openFDA)

Terumo Cardiovascular Systems Corporation: Occluder module, catalog 803480, Terumo Cardiovascular System Corp, Ann Arbor, MI. The occluder module is used with the Occluder head to provide a computer controlled tube clamping mechanism to regulate the flow in the venous line.

Recall number
Z-0883-2011
Recalling firm
Terumo Cardiovascular Systems Corporation
Product
Occluder module, catalog 803480, Terumo Cardiovascular System Corp, Ann Arbor, MI. The occluder module is used with the Occluder head to provide a computer controlled tube clamping mechanism to regulate the flow in the venous line.
Status
Terminated
Initiated
2010-12-02
Posted
2011-01-07
Terminated
2012-04-17
Reason
The firm identified that a component within the System 1 ABD and Occluder Modules that may malfunction. The Air Bubble Detector Module (ABD) is used with an air sensor to detect gross air bubbles in the perfusion circuit. The Occluder module is used with the Occluder head to provide a computer controlled tube clamping mechanism to regulate the flow in the venous line. The firm identified a fai
Root cause
Component design/selection
Action
Terumo Cardiovascular Systems Corporation issued an URGENT MEDICAL DEVICE CORRECTION letter dated 12/2/2010 to the Director of Cardiac Operating Room Services or Operating Room Nurse Manager. The letter identified the product, the problem, and the action to be taken. The letter stated the firm will replace all modules in the affected population, and that a representative would contact the consignee regarding the replacement schedule. Customers were to review the Medical Device Correction. Assure that all user were aware of the notice. Confirm receipt of the communication by faxing the attached Customer Response Form to the fax number indicated on the form. For questions regarding this recall call 1-800-521-2818.
Quantity
221 devices
Distribution
Worldwide Distribution - US, Austrailia, Belgium, Canada, Chile, Columbia, Costa Rica, Dominican Republic, Guatamala, Hong Kong, Indonesia, Japan, Korea, Malaysia, Mexico, Phillipines, Saudi Arabia, Singapore, Taiwan, Thailand, United Arab Emirates, and Vietnam
Lot, serial or model codes
00632-00635, 00638, 00642-00711, 00713-00715, 00717-00741, 00752-00813, 00815-00821, 00823-00864, 00867, 00869-00874.
510(k) numbers
not on file
PMA numbers
not on file
Product code
DTQ
Firm FEI
1828100
Firm city
Ann Arbor
Firm state
MI
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register