FDA Medical Device Recalls (openFDA)
Arjo, Inc.: Minstrel Patient Lift; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model numbers HMB001-US and HMB002-US (with scale)
- Recall number
- Z-0883-04
- Recalling firm
- Arjo, Inc.
- Product
- Minstrel Patient Lift; a non-AC-powered patient lift; Arjo, Inc., 50 N. Gary Ave., Roselle, IL 60172; model numbers HMB001-US and HMB002-US (with scale)
- Status
- Terminated
- Initiated
- 2003-01-30
- Posted
- 2004-07-20
- Terminated
- 2004-05-12
- Reason
- The tension pin was not installed in spreader bar pivot bolt to prevent the pivot bolt from migrating out of the spreader bar, allowing the spreader bar to detach from the lift.
- Root cause
- Other
- Action
- All affected end users were given a copy of the Customer Advisory Notice dated 1/30/03, during the Arjo salesmen visits to inspect the Minstrel lifts for the presence of the tension pin in the pivot bolt of the lift. The advisory informed the user of the incident where the spreader bar detached from the scale unit. The end user was requested to read the advisory and sign the bottom portion of it certifying that they read the advisory and that the Arjo salesman inspected their units for the presence of the tension pin, retaining a copy of the signed advisory notice and the inspeciton record.
- Quantity
- 46 units
- Distribution
- Nationwide. District of Columbia, New York, Washington, Iowa, Florida, North Carolina, Louisiana, Virginia, South Carolina, Minnesota and Colorado
- Lot, serial or model codes
- model numbers HMB001-US and HMB002-US, all units with and without scales up to and including serial number MPLA0203A0464
- 510(k) numbers
- not on file
- PMA numbers
- not on file
- Product code
- FSA
- Firm FEI
- 1419652
- Firm city
- Roselle
- Firm state
- IL
- Firm country
- not on file
Where this comes from
- Register
- FDA Medical Device Recalls (openFDA)
- Publisher
- U.S. Food and Drug Administration (openFDA)
- File
- device-recall-0001-of-0001.json.zip
- sha256
- b87518bfe2e8
- Size
- 288210821 bytes
- Published
- 2026-09-28
- Read
- 2026-09-29T08:25:23.626Z
- Source file
- Source file retained · verify
- Refresh due
- 2026-10-06
Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.
Cite as
FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28