Skip to the content

FDA Medical Device Recalls (openFDA)

Biomet, Inc.: A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.

Recall number
Z-0882-2021
Recalling firm
Biomet, Inc.
Product
A.L.P.S. Clavicle Plating System, Plate, Fixation, Bone, Item Number 233500004 - Product Usage: indicated for fixation of fractures, osteotomies and non-unions of the clavicle including osteopenic bone.
Status
Terminated
Initiated
2020-12-23
Posted
2021-01-21
Terminated
2023-04-13
Reason
The instruments may become cold welded together when the drill is inserted through the Soft Tissue Guide. This prevents disassociation of the instruments and potentially leads to an increased chance of fracture of the drill.
Root cause
Device Design
Action
On 12/30/2020, Zimmer Biomet issued an Urgent Medical Device Recall to the consumer/user level, including any intermediate wholesale or retail consignees. All distributors will be notified via email. Hospital risk managers and surgeons, as well as distributors with product, will be notified via courier.
Quantity
174 systems
Distribution
Nationwide distribution.
Lot, serial or model codes
Item: 233500004; Lot: 516349, 545167; UDI: (01)0088030483464(10)516349, (01)00880304834644(11)190208(10)545167
510(k) numbers
["K173767"]
PMA numbers
not on file
Product code
HRS
Firm FEI
1825034
Firm city
Warsaw
Firm state
IN
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register