Skip to the content

FDA Medical Device Recalls (openFDA)

Amendia, Inc: K-Wire 1.4mm x 12in Non-Threaded, Part Number 9080-12U; K-Wire Threaded 1.4mm x 18, Part Number 9080-18T; K-Wire Non-Threaded 1.4mm x 18, Part Number 9080-18U; 1.4mm x 24 Threaded Guide Wire, Part Number 9080-24T; K-Wire Non-Threaded 24, Part Number 9080-24U; Blunt K-Wire, 18 inches, Threaded, Part Number 9080B-18T; Blunt K-Wire, 18 inches, Unthreaded, Part Number 9080B-18U; Blunt K-Wire, 24 inches, Unthreaded, Part Number 9080B-24U; 1.4mm x 18 Blunt Threaded Nitinol Guide Wire, Part Number 9080B-N-18T; 1.4mm x 18" Blunt Non-Threaded Nitinol Guide Wire, Part Number 9080B-N-18U; Lodestar K-wire Non-Threaded, 18, Part Number 9080L-18U; 1.4mm x 18 Threaded Nitinol Guide Wire, Part Number 9080-N-18T; 1.4mm x 18 Non-Threaded Nitinol Guide Wire, Part Number 9080-N-18U The guide wire is passed through the cannulated handle of the Dual Stylet Needles and pressed into the bone

Recall number
Z-0881-2017
Recalling firm
Amendia, Inc
Product
K-Wire 1.4mm x 12in Non-Threaded, Part Number 9080-12U; K-Wire Threaded 1.4mm x 18, Part Number 9080-18T; K-Wire Non-Threaded 1.4mm x 18, Part Number 9080-18U; 1.4mm x 24 Threaded Guide Wire, Part Number 9080-24T; K-Wire Non-Threaded 24, Part Number 9080-24U; Blunt K-Wire, 18 inches, Threaded, Part Number 9080B-18T; Blunt K-Wire, 18 inches, Unthreaded, Part Number 9080B-18U; Blunt K-Wire, 24 inches, Unthreaded, Part Number 9080B-24U; 1.4mm x 18 Blunt Threaded Nitinol Guide Wire, Part Number 9080B-N-18T; 1.4mm x 18" Blunt Non-Threaded Nitinol Guide Wire, Part Number 9080B-N-18U; Lodestar K-wire Non-Threaded, 18, Part Number 9080L-18U; 1.4mm x 18 Threaded Nitinol Guide Wire, Part Number 9080-N-18T; 1.4mm x 18 Non-Threaded Nitinol Guide Wire, Part Number 9080-N-18U The guide wire is passed through the cannulated handle of the Dual Stylet Needles and pressed into the bone
Status
Terminated
Initiated
2016-11-22
Posted
not on file
Terminated
2024-02-01
Reason
The rods are packaged in packages with unacceptable seals. In some cases, the integrity of the seal is compromised by channels or voids.
Root cause
Employee error
Action
Amendia sent an Urgent Medical Device Recall letter dated November 22, 2016, to all affected customers. The letter requested that customers examine their inventory and quarantine the product. The letter also requested that they contact Amendia to confirm receipt of the letter and to obtain a Return Material Authorization for any product on hand. Customers with questions were instructed to call the following numbers: 770-575-5224 770-575-5236 770-575-5369 For questions regarding this recall call 404-610-7215.
Quantity
5833 units
Distribution
Nationwide Distribution to AL, AZ, AR, CA, CO, FL, GA, IA, KS, KY, LA, MA, MI, MN, MO, NV, NJ, NC, SC, TN, TX, VA, WA, WV, WI and Puerto Rico
Lot, serial or model codes
Part Number 9080-12U Lot Number 121306, 121307, 122628, 122721, 124400, 132732, 135403; Part Number 9080-18T Lot Number 122486, 122626, 123957, 124408, 127435, 129132, 129803, 129935, 131072, 135560, 135566, 135567; Part Number 9080-18U Lot Number 122470, 123948, 129802, 129936, 130390, 130391, 131073; Part Number 9080-24T Lot Number 133359; Part Number 9080-24U Lot Number 134108, 134112, 134908; Part Number 9080B-18T Lot Number 124547, 126820, 127437, 131454, 132592, 134181, 135401; Part Number 9080B-18U Lot Number 122468, 123938, 126504, 127436, 129800, 130947, 131453, 132750, 133645, 134169, 134907, 135563, 135578, 135579, 135580; Part Number 9080B-24U Lot Number 122239, 130392, 130393, 131839, 133988, 135359, 135360, 135400; Part Number 9080B-N-18T Lot Number 124137, 125331, 128472, 130612, 132229, 135584, 135585, 137464; Part Number 9080B-N-18U Lot Number 123556, 125199, 126821, 128348, 128469, 129443, 129915, 130836, 131626, 133118, 134180, 135208; Part Number 9080L-18U Lot Number 127877, 128351, 130389, 130946, 133261, 135210, 135593; Part Number 9080-N-18T Lot Number 122488, 123863, 130914; Part Number 9080-N-18U Lot Number 122487, 123564, 124402, 126125, 129005, 129077, 133499, 135562, 137462.
510(k) numbers
not on file
PMA numbers
not on file
Product code
HTY
Firm FEI
3007970248
Firm city
Marietta
Firm state
GA
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register