Skip to the content

FDA Medical Device Recalls (openFDA)

Philips Medical Systems (Cleveland) Inc: Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260. A component of the Computed Tomography X-Ray System.

Recall number
Z-0881-2013
Recalling firm
Philips Medical Systems (Cleveland) Inc
Product
Philips Medical Systems, Extended Brilliance Workspace (EBW) X-Ray Scanners, Model #728260. A component of the Computed Tomography X-Ray System.
Status
Terminated
Initiated
2012-10-24
Posted
2013-02-26
Terminated
2015-07-23
Reason
Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project
Root cause
Software design
Action
Philips Medical issued an URGENT - Medical Device Correction letter to their customers concerning the Tumor LOC IPD issue on October 25, 2012. The letter informs the customer of the nature of this software problem and names the Brilliance and GEMINI scanning systems affected by this issue. The letter goes on to describe the hazard inherent to this software issue and provides 'Workaround procedures which the customers can use in order to help mitigate the effects of this problem. Customers with questions should contact their Customers Care Solutions Center at 1-800-722-9377, option 5. For questions regarding this recall call 1-800-722-9377.
Quantity
39 units
Distribution
Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..
Lot, serial or model codes
The EBW Serial Numbers subject to software updates are as follows: 8222, 8235, 8270, 8838, 8841, 8862, 9014, 9019, 9127, 10331, 10348, 10402, 10705, 10749, 10827, 11002, 11362, 11365, 11588, 11597, 11652, 11693, 11704, 11724, 11725, 11828, 11857, 11861, 11945, 11949, 11950, 11951, 11952, 11953, 11954, 11955, 11956, 11957, and 11979.
510(k) numbers
["K012009"]
PMA numbers
not on file
Product code
JAK
Firm FEI
1525965
Firm city
Cleveland
Firm state
OH
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register