Skip to the content

FDA Medical Device Recalls (openFDA)

Dade Behring , Inc.: Stratus CS STAT Fluorometric Analyzer CKMB DilPak (CCKMB-D) in vitro diagnostic.

Recall number
Z-0881-06
Recalling firm
Dade Behring , Inc.
Product
Stratus CS STAT Fluorometric Analyzer CKMB DilPak (CCKMB-D) in vitro diagnostic.
Status
Terminated
Initiated
2006-04-03
Posted
2006-05-19
Terminated
2006-08-07
Reason
Foil sealing of the DilPak may occlude the pipet tip resulting in insufficient aspiration of fluids. Erroneous test results may be generated without an associated error code. Test results may be falsely elevated or depressed, and the magnitude of the inaccuracy may vary based on the degree of occlusion.
Root cause
Other
Action
Customers were provided clear written instructions (04/06/2006) to discontinue use and discard remaining inventory of products over 2 months old. In addition, each customer is being contacted by phone as an effectiveness verification. Recalling firm is also instructing Stratus CS customers to process all samples and dilutions in duplicate.
Quantity
246 cartons of 5 DilPaks
Distribution
Nationwide to hospitals and government accounts in addition to International accounts in Canada, Saudi Arabia, Bangladesh, UAE, India, Japan, Denmark, and Australia.
Lot, serial or model codes
Lot # 845304002 and 846058002
510(k) numbers
["K981099"]
PMA numbers
not on file
Product code
JHX
Firm FEI
2517506
Firm city
Newark
Firm state
DE
Firm country
not on file

Where this comes from

Register
FDA Medical Device Recalls (openFDA)
Publisher
U.S. Food and Drug Administration (openFDA)
File
device-recall-0001-of-0001.json.zip
sha256
b87518bfe2e8
Size
288210821 bytes
Published
2026-09-28
Read
2026-09-29T08:25:23.626Z
Source file
Source file retained · verify
Refresh due
2026-10-06

Unless otherwise noted, the content, data, documentation, code, and related materials on openFDA is public domain and made available with a Creative Commons CC0 1.0 Universal dedication.

Cite as

FDA Medical Device Recalls (openFDA) (U.S. Food and Drug Administration (openFDA)), device-recall-0001-of-0001.json.zip, sha256 b87518bfe2e8, published 2026-09-28

Search this register